J Biomed Opt. 2017 Sep;22(9):1-9. doi: 10.1117/1.JBO.22.9.097002.
The response of a fiber optic sensor [linearly chirped fiber Bragg grating (LCFBG)] to a linear thermal gradient applied on its sensing length (i.e., 1.5 cm) has been investigated. After these bench tests, we assessed their feasibility for temperature monitoring during thermal tumor treatment. In particular, we performed experiments during ex vivo laser ablation (LA) in pig liver and in vivo thermal ablation in animal models (pigs). We investigated the following: (i) the relationship between the full width at half maximum of the LCFBG spectrum and the temperature difference among the extremities of the LCFBG and (ii) the relationship between the mean spectrum wavelength and the mean temperature acting on the LCFBG sensing area. These relationships showed a linear trend during both bench tests and LA in animal models. Thermal sensitivity was significant although different values were found with regards to bench tests and animal experiments. The linear trend and significant sensitivity allow hypothesizing a future use of this kind of sensor to monitor both temperature gradient and mean temperature within a tissue undergoing thermal treatment.
研究了一种光纤传感器[线性啁啾光纤布拉格光栅(LCFBG)]对施加在其传感长度(即1.5厘米)上的线性热梯度的响应。经过这些台架试验后,我们评估了它们在热肿瘤治疗期间进行温度监测的可行性。具体而言,我们在猪肝的离体激光消融(LA)以及动物模型(猪)的体内热消融过程中进行了实验。我们研究了以下内容:(i)LCFBG光谱半高宽与LCFBG两端温度差之间的关系,以及(ii)平均光谱波长与作用在LCFBG传感区域的平均温度之间的关系。在台架试验和动物模型的LA过程中,这些关系均呈现出线性趋势。尽管在台架试验和动物实验中发现了不同的热灵敏度值,但热灵敏度是显著的。这种线性趋势和显著的灵敏度使得我们可以推测,未来这种类型的传感器可用于监测正在接受热处理的组织内的温度梯度和平均温度。