Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital.

作者信息

Akbas Neval, Gonzalez Graciela, Edwards Rachel, Devaraj Sridevi

机构信息

Department of Pathology & Immunology, Baylor College of Medicine, 6621 Fannin Street, Houston, Tx 77030, USA.

Texas Children's Hospital, 6621 Fannin Street, Houston, Tx 77030, USA.

出版信息

Pract Lab Med. 2015 Oct 3;3:1-7. doi: 10.1016/j.plabm.2015.09.002. eCollection 2015 Dec 1.

Abstract

OBJECTIVES

Point of care testing (POCT) contributes to diagnosis and monitoring with fast testing time and easily performed assays. We evaluated the Abaxis Piccolo Xpress point of care chemistry analyzer using the Liver Panel Plus discs for our pediatric patient population at Texas Children's Hospital.

DESIGN AND METHODS

Analytical performance was evaluated for precision and linearity using quality control materials and commercially available verification samples. Comparison studies were performed between Piccolo Xpress analyzer and Vitros 5600 analyzer using patient samples. Interference studies were carried out using nine different patient pool sera. Lipemia interference was removed using LipoClear for severely lipemic sample pools.

RESULTS

Precision of all tests was excellent (CVs<5% for all measured analytes except TBIL). All assays were linear and accurate within the allowable total error. Comparison studies showed that three analytes (amylase, GGT and TBIL) had statistically significant bias. Interference study results did not exceed the total allowable error for hemoglobin (<150 mg/dL), bilirubin (<15 mg/dL) and lipemia (<400 mg/dL except ALT, GGT and TP). LipoClear treatment removed lipemia interference for all analytes except total protein.

CONCLUSIONS

The Piccolo Xpress chemistry analyzer showed an acceptable analytical performance for precision, linearity and interference from common substances. Increased bias for three analytes in comparison studies could be due to different methodologies.

摘要

目的

即时检验(POCT)凭借快速的检测时间和易于操作的检测方法,有助于疾病的诊断和监测。我们在德克萨斯儿童医院,针对儿科患者群体,使用肝脏检测组合 Plus 试剂盘,对 Abaxis Piccolo Xpress 即时化学分析仪进行了评估。

设计与方法

使用质量控制材料和市售验证样本,对分析性能的精密度和线性进行评估。使用患者样本,在 Piccolo Xpress 分析仪和 Vitros 5600 分析仪之间进行比较研究。使用九种不同的患者混合血清进行干扰研究。对于严重脂血样本池,使用 LipoClear 去除脂血干扰。

结果

所有检测的精密度都非常出色(除总胆红素外,所有被测分析物的变异系数<5%)。所有检测方法在允许的总误差范围内呈线性且准确。比较研究表明,三种分析物(淀粉酶、γ-谷氨酰转移酶和总胆红素)存在统计学上的显著偏差。干扰研究结果未超过血红蛋白(<150mg/dL)、胆红素(<15mg/dL)和脂血(除谷丙转氨酶、γ-谷氨酰转移酶和总蛋白外<400mg/dL)的总允许误差。LipoClear 处理消除了除总蛋白外所有分析物的脂血干扰。

结论

Piccolo Xpress 化学分析仪在精密度、线性以及常见物质干扰方面表现出可接受的分析性能。比较研究中三种分析物偏差增加可能是由于方法不同所致。

相似文献

1
Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital.
Pract Lab Med. 2015 Oct 3;3:1-7. doi: 10.1016/j.plabm.2015.09.002. eCollection 2015 Dec 1.
2
Liver function testing on the Abaxis Piccolo Xpress: Use in Ebola virus disease protocols.
Clin Chim Acta. 2015 Jun 15;446:119-27. doi: 10.1016/j.cca.2015.03.044. Epub 2015 Apr 16.
3
Analytical performance of the Abaxis Piccolo Xpress® point of care analyzer in whole blood, serum, and plasma.
Clin Biochem. 2015 Dec;48(18):1344-6. doi: 10.1016/j.clinbiochem.2015.08.002. Epub 2015 Aug 6.
4
[Performance evaluation of the Piccolo xpress Point-of-care Chemistry Analyzer].
Korean J Lab Med. 2009 Oct;29(5):430-8. doi: 10.3343/kjlm.2009.29.5.430.
5
Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood, serum, and plasma.
Data Brief. 2017 Nov 4;16:81-89. doi: 10.1016/j.dib.2017.11.006. eCollection 2018 Feb.
7
Effect of interference from hemolysis, icterus and lipemia on routine pediatric clinical chemistry assays.
Clin Chim Acta. 2015 Jan 1;438:241-5. doi: 10.1016/j.cca.2014.08.008. Epub 2014 Aug 13.
8
Effects of marked hypertriglyceridemia and lipid clearance techniques on canine biochemistry testing.
J Vet Diagn Invest. 2019 Jul;31(4):546-553. doi: 10.1177/1040638719858690. Epub 2019 Jun 25.
9
The evaluation and comparison of consecutive high speed centrifugation and LipoClear® reagent for lipemia removal.
Clin Biochem. 2014 Mar;47(4-5):309-14. doi: 10.1016/j.clinbiochem.2014.01.001. Epub 2014 Jan 14.
10
Evaluation of the Nova StatSensor® Xpress(TM) Creatinine point-of-care handheld analyzer.
PLoS One. 2015 Apr 17;10(4):e0122433. doi: 10.1371/journal.pone.0122433. eCollection 2015.

本文引用的文献

1
Liver function testing on the Abaxis Piccolo Xpress: Use in Ebola virus disease protocols.
Clin Chim Acta. 2015 Jun 15;446:119-27. doi: 10.1016/j.cca.2015.03.044. Epub 2015 Apr 16.
2
A point-of-care chemistry test for reduction of turnaround and clinical decision time.
Am J Emerg Med. 2011 Jun;29(5):489-95. doi: 10.1016/j.ajem.2009.11.020. Epub 2010 Apr 24.
3
[Performance evaluation of the Piccolo xpress Point-of-care Chemistry Analyzer].
Korean J Lab Med. 2009 Oct;29(5):430-8. doi: 10.3343/kjlm.2009.29.5.430.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验