Cheng H, Pittman K A, Dandekar K A
Metabolism and Pharmacokinetics Department, Bristol-Myers Company, Buffalo, NY 14213.
J Pharm Sci. 1987 Dec;76(12):918-9. doi: 10.1002/jps.2600761216.
A simple and sensitive assay for quantitating 9-methyl-3-(1H-tetrazol-5-yl)-4H-pyrido[1,2-a]pyrimidin-4-one (1; BMY 26517) in human plasma was developed using high-performance liquid chromatography with fluorescence detection. The method involves precipitation of protein and reversed-phase chromatography. The method is linear in the range of 4.3-429 ng/mL of 1, and the limit of detection is 0.4 ng/mL. The day-to-day precision values of this method at 25.7 and 386 ng/mL are 2.1 and 2.6%, respectively. The day-to-day accuracy values at these concentrations are 99.7 and 99.8%, respectively. The recovery of 1 is 98.3%.