Clinical evaluation of a prosthetic suspension system: Looped silicone liner.

作者信息

Abu Osman Noor Azuan, Gholizadeh Hossein, Eshraghi Arezoo, Wan Abas Wan Abu Bakar

机构信息

1 BioApps Sdn Bhd, PlatCOM Ventures Sdn Bhd, Centre for Applied Biomechanics, Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia.

2 Ottawa Hospital Research Institute, Centre for Rehabilitation Research and Development, Ottawa, Canada.

出版信息

Prosthet Orthot Int. 2017 Oct;41(5):476-483. doi: 10.1177/0309364616670396. Epub 2016 Oct 6.

Abstract

OBJECTIVES

This study aimed to evaluate and compare a newly designed suspension system with a common suspension in the market.

STUDY DESIGN

Prospective study.

METHODS

Looped liners with hook fastener and Iceross Dermo Liner with pin/lock system were mechanically tested using a tensile testing machine in terms of system safety. A total of 10 transtibial amputees participated in this study and were asked to use these two different suspension systems. The pistoning was measured between the liner and socket through a photographic method. Three static axial loading conditions were implemented, namely, 30, 60, and 90 N. Furthermore, subjective feedback was obtained.

RESULTS

Tensile test results showed that both systems could safely tolerate the load applied to the prosthesis during ambulation. Clinical evaluation confirmed extremely low pistoning in both systems (i.e. less than 0.4 cm after adding 90 N traction load to the prosthesis). Subjective feedback also showed satisfaction with both systems. However, less traction at the end of the residual limb was reported while looped liner was used.

CONCLUSION

The looped liner with hook fastener is safe and a good alternative for individuals with transtibial amputation as this system could solve some problems with the current systems. Clinical relevance The looped liner and hook fastener were shown to be good alternative suspension for people with lower limb amputation especially those who have difficulty to use and align the pin/lock systems. This system could safely tolerate centrifugal forces applied to the prosthesis during normal and fast walking.

摘要

目的

本研究旨在评估并比较一种新设计的悬吊系统与市场上常见的悬吊系统。

研究设计

前瞻性研究。

方法

使用拉伸试验机对带有钩扣的环形衬垫和带有销钉/锁定系统的Iceross Dermo衬垫进行系统安全性方面的机械测试。共有10名经胫截肢者参与本研究,并被要求使用这两种不同的悬吊系统。通过摄影方法测量衬垫与接受腔之间的活塞运动情况。实施了三种静态轴向加载条件,即30 N、60 N和90 N。此外,还获得了主观反馈。

结果

拉伸测试结果表明,两种系统在步行过程中都能安全承受施加在假肢上的负荷。临床评估证实两种系统中的活塞运动都极低(即在给假肢施加90 N牵引负荷后小于0.4厘米)。主观反馈也表明对两种系统都感到满意。然而,在使用环形衬垫时,残肢末端的牵引力较小。

结论

带有钩扣的环形衬垫是安全的,对于经胫截肢者来说是一个很好的选择,因为该系统可以解决当前系统存在的一些问题。临床意义 环形衬垫和钩扣被证明是下肢截肢者的良好替代悬吊装置,尤其是那些在使用和调整销钉/锁定系统方面有困难的人。该系统在正常和快速行走过程中能够安全承受施加在假肢上的离心力。