A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions.

作者信息

Langton D J, Sidaginamale R P, Joyce T J, Meek R D, Bowsher J G, Deehan D, Nargol A V F, Holland J P

机构信息

Freeman Hospital, Northern Retrieval Registry, University Hospital of North Tees and Freeman Hospital, Newcastle Upon Tyne, UK.

Newcastle University, Stephenson Building, Newcastle Upon Tyne, NE1 7RU, UK.

出版信息

Bone Joint J. 2017 Oct;99-B(10):1304-1312. doi: 10.1302/0301-620X.99B10.BJJ-2016-1005.R1.

Abstract

AIMS

We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.

PATIENTS AND METHODS

We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.

RESULTS

After adjustment for other factors, CoCr stem tapers were found to have significantly greater volumetric material loss than the equivalent Ti stem tapers.

CONCLUSION

When taper junction damage is identified during revision of a LD MoM hip, it should be suspected that a male taper composed of a standard CoCr alloy has sustained significant changes to the taper cone geometry which are likely to be more extensive than those affecting a Ti alloy stem. Cite this article: 2017;99-B:1304-12.

摘要

目的

我们试图确定在大直径金属对金属(MoM)髋关节置换系统中使用时,钴铬合金(CoCr)股骨柄锥度(耳轴)的磨损是否比钛(Ti)合金柄锥度(耳轴)更严重。

患者与方法

我们采用经过验证的方法进行取出分析,以确定取出的大直径CoCr MoM髋关节置换术中,与钛合金股骨柄(n = 28)或CoCr股骨柄(n = 21)一起使用时,锥度表面的体积材料损失。仅纳入了14个锥度结构中的12个,其粗糙的阳锥度表面且标称夹角接近5.666°。使用锥度角、锥度粗糙度、轴承直径(水平杠杆臂)作为自变量进行多元回归建模。使用坐标测量机、轮廓仪和扫描电子显微镜对材料损失进行测绘。

结果

在对其他因素进行调整后,发现CoCr柄锥度的体积材料损失明显大于同等的Ti柄锥度。

结论

在大直径MoM髋关节翻修术中发现锥度连接损伤时,应怀疑由标准CoCr合金制成的阳锥度对锥度圆锥几何形状产生了显著变化,这些变化可能比影响钛合金柄的变化更广泛。引用本文:2017;99 - B:1304 - 12。