Abstract

INTRODUCTION

Rotational Thromboelastometry (ROTEM) is a point of care method used to monitor coagulation during surgery and to guide transfusion strategies in patients presenting with severe bleeding. The aim of our study was to determine the impact of four direct oral anticoagulants (DOACs) on 3 commonly used ROTEM tests.

METHODS

Whole blood samples from 20 healthy donors were spiked in vitro with apixaban, edoxaban, rivaroxaban or dabigatran at 5 different plasma concentrations (0-1000 ng/mL). EXTEM, INTEM and FIBTEM tests were systematically performed.

RESULTS

There was a linear relationship between the increase in clotting time (CT) and plasma DOAC concentrations in both the EXTEM and INTEM tests. We found that the DOAC concentration required to double EXTEM CT was 1042 ± 225 ng/mL for apixaban, 134 ± 38 ng/mL for edoxaban, 176 ± 26 ng/mL for rivaroxaban and 284 ± 73 ng/mL for dabigatran. INTEM CT was less sensitive than EXTEM CT whatever the anticoagulant. EXTEM CT was above the normal range for 5 of 5 spiked samples when the plasma concentrations were ~1000 ng/mL for apixaban, ~100 ng/mL for edoxaban, ~200 ng/mL for rivaroxaban and ~200 ng/mL for dabigatran. Maximum Clot Firmness in EXTEM, INTEM and FIBTEM tests was not affected whatever the DOAC or its concentration.

CONCLUSION

This study found a DOAC dose-dependent increase in ROTEM CTs. ROTEM tests were only poorly impacted by low levels of edoxaban, rivaroxaban or dabigatran. Apixaban had only a low effect even at high concentrations.

摘要

简介

旋转血栓弹性测定法(ROTEM)是一种用于监测手术期间凝血的即时检测方法,用于指导出现严重出血的患者的输血策略。我们的研究目的是确定四种直接口服抗凝剂(DOAC)对三种常用 ROTEM 检测的影响。

方法

从 20 名健康供体中采集全血样本,在体外用阿哌沙班、依度沙班、利伐沙班或达比加群以 5 种不同的血浆浓度(0-1000ng/mL)进行预混。系统地进行 EXTEM、INTEM 和 FIBTEM 检测。

结果

在 EXTEM 和 INTEM 检测中,凝血时间(CT)的增加与血浆 DOAC 浓度呈线性关系。我们发现,使 EXTEM CT 加倍所需的 DOAC 浓度为阿哌沙班 1042±225ng/mL,依度沙班 134±38ng/mL,利伐沙班 176±26ng/mL,达比加群 284±73ng/mL。无论抗凝剂如何,INTEM CT 都不如 EXTEM CT 敏感。当血浆浓度约为 1000ng/mL 时,阿哌沙班 5 个样本中有 5 个样本的 EXTEM CT 超过正常范围,当血浆浓度约为 100ng/mL 时,依度沙班 5 个样本中有 5 个样本的 EXTEM CT 超过正常范围,当血浆浓度约为 200ng/mL 时,利伐沙班 5 个样本中有 5 个样本的 EXTEM CT 超过正常范围,当血浆浓度约为 200ng/mL 时,达比加群 5 个样本中有 5 个样本的 EXTEM CT 超过正常范围。EXTEM、INTEM 和 FIBTEM 检测中的最大凝块硬度不受 DOAC 或其浓度的影响。

结论

本研究发现 ROTEM CT 随 DOAC 剂量增加呈剂量依赖性增加。即使在低浓度下,EDOX、RIVAR 和 DABI 对 ROTEM 检测的影响也很小。即使在高浓度下,阿哌沙班的作用也很低。