Pedretti Stefano, Cipriani Manlio, Bonacina Edgardo, Vargiu Sara, Gil Ad Vered, Frigerio Maria, Lunati Maurizio
Electrophysiology Unit, "De Gasperis" Cardio Center, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy.
Heart Failure Unit, "De Gasperis" Cardio Center, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy.
J Arrhythm. 2017 Oct;33(5):494-496. doi: 10.1016/j.joa.2017.04.007. Epub 2017 May 11.
In patients with left ventricular assist device (LVAD), a minority of post-operative ventricular tachycardias (VTs) is caused by contact between the inflow cannula and the endocardium. Currently, electrophysiologic characteristics and pathologic features of this condition are lacking. We report on a case of a successfully ablated mechanical VT. After VT recurrence, heart transplantation took place. Pathologic observations were consistent with direct tissue injury and inflammation, eventually contributing to persisting arrhythmias. Radiofrequency catheter ablation can be a safe and effective option to treat arrhythmias caused by inflow cannula interference in the short term, although a high recurrence rate is expected.
在植入左心室辅助装置(LVAD)的患者中,少数术后室性心动过速(VT)是由流入插管与心内膜接触引起的。目前,关于这种情况的电生理特征和病理特征尚不明确。我们报告了一例成功消融的机械性VT病例。VT复发后,进行了心脏移植。病理观察结果与直接组织损伤和炎症一致,最终导致心律失常持续存在。射频导管消融在短期内可以是治疗由流入插管干扰引起的心律失常的一种安全有效的选择,尽管预期复发率较高。