Andrews B T, Levy M, McIntosh T K, Pitts L H
Department of Neurological Surgery, School of Medicine, University of California, San Francisco 94143.
Neurol Res. 1988 Jun;10(2):123-6. doi: 10.1080/01616412.1988.11739828.
We described the construction and use of a simple and reliable catheter system that can be used to monitor intracranial pressure (ICP) from the epidural space in rats in an experimental setting. The catheter system is easily fabricated in the laboratory from readily available materials. The monitor is fitted flush with the inner table through a burr hole in the temporal squama. A side port is used to fill the system with saline and to irrigate the system should be catheter become obstructed. The distal end of the catheter is fitted to a pressure transducer that is connected to a graphic display and recording system. This system was used to record ICP in 30 anaesthetized adult rats. Seven were subjected to baseline ICP recording only and 23 were subjected to baseline recordings followed by epidural balloon compression of the contralateral hemisphere. Baseline ICP varied between 0 and 8 mmHg and respiratory variation could be detected on the tracings for 24 rats (75%). ICP responded directly and sensitively to epidural balloon inflation in all 23 rats tested. In 5 rats that died during balloon inflation, the decrease in ICP after death followed closely the loss of arterial blood pressure. There was a close correlation of numerical values obtained in two rats in which ICP was recorded simultaneously from the epidural catheter and from a catheter in the subarachnoid space at the cisterna magna. In one rat in which ICP increased to more than 70 mmHg, the epidural catheter continued to record ICP accurately while the cisternal catheter became obstructed with herniated brain.(ABSTRACT TRUNCATED AT 250 WORDS)
我们描述了一种简单可靠的导管系统的构建与使用方法,该系统可用于在实验环境中监测大鼠硬膜外间隙的颅内压(ICP)。该导管系统易于在实验室用现成材料制作。监测器通过颞鳞部的骨孔与内板齐平安装。一个侧端口用于向系统中注入生理盐水,并且在导管堵塞时对系统进行冲洗。导管的远端连接到一个压力传感器,该传感器与图形显示和记录系统相连。此系统用于记录30只麻醉成年大鼠的ICP。7只仅进行基线ICP记录,23只先进行基线记录,随后对侧半球进行硬膜外球囊压迫。基线ICP在0至8 mmHg之间变化,24只大鼠(75%)的记录曲线上可检测到呼吸变化。在所有23只受试大鼠中,ICP对硬膜外球囊充气有直接且敏感的反应。在球囊充气过程中死亡的5只大鼠中,死亡后ICP的下降与动脉血压的丧失密切相关。在两只大鼠中,分别从硬膜外导管和枕大池蛛网膜下腔导管同时记录ICP,所得数值密切相关。在一只ICP升高至超过70 mmHg的大鼠中,硬膜外导管继续准确记录ICP,而枕大池导管因脑疝而堵塞。(摘要截短于250字)