Amoo Olufemi S, Taiwo Idowu A, Salu Olumuyiwa O, Okwuraiwe Azuka P, Onwuamah Chika K, Awe Morenike A, Oforomeh Osaga O, Onwujekwe Daniel I, Ezechi Oliver C, Ajuma Audu R
Human Virology Laboratory, Nigerian Institute of Medical Research, Yaba, Lagos, Nigeria.
Department of Cell Biology & Genetics, University of Lagos, Akoka, Lagos, Nigeria.
Afr J Lab Med. 2013 May 15;2(1):68. doi: 10.4102/ajlm.v2i1.68. eCollection 2013.
The use of real-time Polymerase chain reaction (PCR) technology options is increasing in resource-limited settings because they are faster, improve assay sensitivity, have higher throughput, larger dynamic ranges and reduced rates of contamination. In 2010, UNAIDS ranked Nigeria as the second highest population of people living with HIV and AIDS (2.98 million people) in the world.
The objective of this study was to compare the analytical performances of the Amplicor HIV-1 Monitor (version 1.5) and the COBAS Ampliprep/Taqman (version 2.0) used in monitoring HIV disease progression in HIV-infected individuals.
In a cross-sectional study, HIV-1 RNA values obtained with the Amplicor HIV-1 monitor version 1.5 were compared with those of the COBAS/Ampliprep TaqMan HIV-1 version 2.0 in a routine clinical setting. Between May and November 2011, 176 plasma samples collected were analysed in parallel using both techniques. Data analysis was done using statgraphics Centurion XVI and Medcalc version 12.0.
The correlation coefficient for the two assays was 0.83 and the level of agreement using a Bland-Altman plot was 94.2%.
These findings suggest that the results from the two methods were comparable, hence the COBAS/Ampliprep Taqman version 2.0 is recommended for high-volume laboratories.
在资源有限的环境中,实时聚合酶链反应(PCR)技术的使用越来越多,因为它们速度更快、提高了检测灵敏度、具有更高的通量、更大的动态范围且污染率更低。2010年,联合国艾滋病规划署将尼日利亚列为世界上艾滋病毒和艾滋病感染者人数第二多的国家(298万人)。
本研究的目的是比较用于监测艾滋病毒感染者疾病进展的Amplicor HIV-1监测仪(版本1.5)和COBAS Ampliprep/Taqman(版本2.0)的分析性能。
在一项横断面研究中,在常规临床环境中,将使用Amplicor HIV-1监测仪版本1.5获得的HIV-1 RNA值与COBAS/Ampliprep TaqMan HIV-1版本2.0的HIV-1 RNA值进行比较。2011年5月至11月期间,使用这两种技术对收集的176份血浆样本进行了平行分析。使用Statgraphics Centurion XVI和Medcalc版本12.0进行数据分析。
两种检测方法的相关系数为0.83,使用Bland-Altman图的一致性水平为94.2%。
这些发现表明两种方法的结果具有可比性,因此推荐COBAS/Ampliprep Taqman版本2.0用于大容量实验室。