Calibration and data extraction in nonoptimized Mueller matrix polarimeters.

作者信息

Rodríguez-Nuñez Omar, López-Téllez Juan Manuel, Rodríguez-Herrera Oscar G, Bruce Neil C

出版信息

Appl Opt. 2017 May 20;56(15):4398-4405. doi: 10.1364/AO.56.004398.

Abstract

We present a method for calibration and data extraction for a nonoptimized Mueller matrix polarimeter. The advantage of this type of method is a reduction in measurement time for multiwavelength systems or in systems with slow response times. The calibration process requires the measurement of four known polarization devices. Here we use free-space transmission, a horizontal and a vertical linear polarizer, and a quarter-wave retarder with its fast axis at 30° to the horizontal. Experimental measurements of rotating quarter-wave and half-wave retarders show that accurate results can be obtained with the proposed method.

摘要

我们提出了一种用于非优化穆勒矩阵偏振仪的校准和数据提取方法。这类方法的优点是可减少多波长系统或响应时间较慢的系统的测量时间。校准过程需要测量四个已知的偏振器件。在这里,我们使用自由空间传输、一个水平和一个垂直线性偏振器,以及一个快轴与水平方向成30°的四分之一波片。旋转四分之一波片和半波片的实验测量结果表明,该方法能够获得准确的结果。