Hu Peng, Chen Ping, Cao Dawei, Fu Xiaolong, Xiao Chengjian, Wang Heyi
Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China.
Se Pu. 2017 Oct 8;35(10):1023-1027. doi: 10.3724/SP.J.1123.2017.07014.
CPL-1 is a metal organic framework (MOF) with potential application as stationary phase material in gas chromatographic (GC) analysis, due to its large specific surface area, uniform pore size and good quantum sieving effect on hydrogen isotopes at low temperatures. Herein, a microporous column packed with CPL-1 was used at cryogenic temperature (77 K), and the column was 1.0 mm in inner diameter, 0.5 m in length. Single crystals of AlO was used to build flow path for chromatographic carrier gas. The results showed that the adsorption of H and D with CPL-1 was 4 mmol/g at 77 K, which was better than MnCl/ -AlO and -AlO. With the injection volume increasing from 0.25 mL to 2 mL, the results showed good linear relationship, and the relative error was less than 4%. The results indicated that the column packed with CPL-1 has wide linear range, good repeatability and high accuracy, and it has potential use in hydrogen isotope analysis with gas chromatography.