Chen Ping, Fu Xiaolong, Hu Peng, Xiao Chengjian, Ren Xingbi, Xia Xiulong, Wang Heyi
Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, Mianyang 621900, China.
Se Pu. 2017 Jul 8;35(7):766-771. doi: 10.3724/SP.J.1123.2017.02028.
The conventional packed column gas chromatographic analysis of hydrogen isotopes has low column efficiency, broad peak and long retention time. In this work, a -AlO with MnCl coated capillary packed column was tested at cryogenic temperature. The systematic column efficiency analysis and the hydrogen isotopes analytical technique research had been carried out. The results showed that, the -AlO with MnCl coating could greatly improve the surface degree of order, pore structure and adsorption properties. Also the -H peak and -H peak were eluted in a single area. The -AlO with MnCl coating was packed into a 0.53 mm inner diameter and 1.0 m long fused silica capillary column. It had a good linear relationship used this column with thermal conductivity detector (TCD) to detect the volume concentrations of hydrogen isotopes from 1 to 10 mL/L, and the relative error was less than 5% for low concentration sample testing. For H, HD and D, the retention times can be shortened to 39, 46 and 60 s, respectively. The limits of detection were reduced to 0.046, 0.067 and 0.072 mL/L, respectively. Compared with conventional packed column, capillary packed column had sharper peak form, higher separation degree of adjacent components, shorter retention time and lower detection limits. The above results indicate that the capillary packed column with TCD detector can be used for fast detection of low concentration of hydrogen isotopes and their online analysis.