Kozlov Mikhail, Horner Marc, Kainz Wolfgang, Angelone Leonardo M
Annu Int Conf IEEE Eng Med Biol Soc. 2017 Jul;2017:1481-1484. doi: 10.1109/EMBC.2017.8037115.
The goal of this work is to investigate the effect of coil losses on the electromagnetic field generated in an ASTM phantom by a birdcage coil. The study was based on different numerical implementations of an RF body coil at 64 MHz, using the same 3D EM and RF circuit co-simulation procedure. The coil quality factor was evaluated with respect to losses due to power feed mismatch and to resistive losses of the coil components. The results of the study showed that the magnetic field at the coil iso-center, normalized to the square root of the whole body specific absorption rate, depends on the coil quality factor.
这项工作的目标是研究线圈损耗对鸟笼式线圈在ASTM体模中产生的电磁场的影响。该研究基于64 MHz射频体线圈的不同数值实现方式,采用相同的三维电磁和射频电路联合仿真程序。针对由于功率馈送失配导致的损耗以及线圈组件的电阻性损耗,对线圈品质因数进行了评估。研究结果表明,在归一化到全身比吸收率平方根的情况下,线圈等中心处的磁场取决于线圈品质因数。