对于无法屏气的患者,二维XD-GRASP比传统二维心脏电影磁共振成像提供更好的图像质量。

Two-dimensional XD-GRASP provides better image quality than conventional 2D cardiac cine MRI for patients who cannot suspend respiration.

作者信息

Piekarski Eve, Chitiboi Teodora, Ramb Rebecca, Latson Larry A, Bhatla Puneet, Feng Li, Axel Leon

机构信息

Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University School of Medicine, 660 First Ave., New York, NY, USA.

Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, 660 First Ave., New York, NY, USA.

出版信息

MAGMA. 2018 Feb;31(1):49-59. doi: 10.1007/s10334-017-0655-7. Epub 2017 Oct 24.

Abstract

OBJECTIVES

Residual respiratory motion degrades image quality in conventional cardiac cine MRI (CCMRI). We evaluated whether a free-breathing (FB) radial imaging CCMRI sequence with compressed sensing reconstruction [extradimensional (e.g. cardiac and respiratory phases) golden-angle radial sparse parallel, or XD-GRASP] could provide better image quality than a conventional Cartesian breath-held (BH) sequence in an unselected population of patients undergoing clinical CCMRI.

MATERIALS AND METHODS

One hundred one patients who underwent BH and FB imaging in a midventricular short-axis plane at a matching location were included. Visual and quantitative image analysis was performed by two blinded experienced readers, using a five-point qualitative scale to score overall image quality and visual signal-to-noise ratio (SNR) grade, with measures of noise and sharpness. End-diastolic and end-systolic left ventricular areas were also measured and compared for both BH and FB images.

RESULTS

Image quality was generally better with the BH cines (overall quality grade for BH vs FB images 4 vs 2.9, p < 0.001; noise 0.06 vs 0.08 p < 0.001; SNR grade 4.1 vs 3, p < 0.001), except for sharpness (p = 0.48). There were no significant differences between BH and FB images regarding end-diastolic or end-systolic areas (p = 0.35 and p = 0.12). Eighteen of the 101 patients had poor BH image quality (grade 1 or 2). In this subgroup, the quality of the FB images was better (p = 0.0032), as was the SNR grade (p = 0.003), but there were no significant differences regarding noise and sharpness (p = 0.45 and p = 0.47).

CONCLUSION

Although FB XD-GRASP CCMRI was visually inferior to conventional BH CCMRI in general, it provided improved image quality in the subgroup of patients with respiratory-motion-induced artifacts on BH images.

摘要

目的

在传统心脏电影磁共振成像(CCMRI)中,残余呼吸运动会降低图像质量。我们评估了一种采用压缩感知重建技术的自由呼吸(FB)径向成像CCMRI序列[超维(如心脏和呼吸相位)金角径向稀疏并行,即XD-GRASP]在接受临床CCMRI检查的未选择患者群体中,是否能提供比传统笛卡尔屏气(BH)序列更好的图像质量。

材料与方法

纳入101例在匹配位置进行心室短轴平面BH和FB成像的患者。由两位经验丰富的盲法阅片者进行视觉和定量图像分析,使用五点定性量表对整体图像质量和视觉信噪比(SNR)等级进行评分,并测量噪声和清晰度。还测量并比较了BH和FB图像的舒张末期和收缩末期左心室面积。

结果

BH电影图像的质量总体上更好(BH与FB图像的整体质量等级分别为4和2.9,p<0.001;噪声分别为0.06和0.08,p<0.001;SNR等级分别为4.1和3,p<0.001),清晰度除外(p=0.48)。BH和FB图像在舒张末期或收缩末期面积方面无显著差异(p=0.35和p=0.12)。101例患者中有18例BH图像质量较差(1级或2级)。在该亚组中,FB图像的质量更好(p=0.0032),SNR等级也是如此(p=0.003),但在噪声和清晰度方面无显著差异(p=0.45和p=0.47)。

结论

虽然总体上FB XD-GRASP CCMRI在视觉上不如传统BH CCMRI,但它在BH图像上因呼吸运动产生伪影的患者亚组中提供了更好的图像质量。

相似文献

本文引用的文献

2
5
Edge Sharpness Assessment by Parametric Modeling: Application to Magnetic Resonance Imaging.基于参数建模的边缘锐度评估:在磁共振成像中的应用
Concepts Magn Reson Part A Bridg Educ Res. 2015 May 1;44(3):138-149. doi: 10.1002/cmr.a.21339. Epub 2015 Sep 28.
6
Real time dynamic MRI by exploiting spatial and temporal sparsity.通过利用空间和时间稀疏性实现实时动态磁共振成像。
Magn Reson Imaging. 2016 May;34(4):473-82. doi: 10.1016/j.mri.2015.10.033. Epub 2015 Nov 11.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验