Investigation of dental cone-beam CT pixel data and a modified method for conversion to Hounsfield unit (HU).

机构信息

Environmental Health & Radiation Safety, University of Pennsylvania, Philadelphia, PA, USA.

DMD Candidate Penn Dental Medicine, University of Pennsylvania, Philadelphia, PA, USA.

出版信息

Dentomaxillofac Radiol. 2018 Feb;47(2):20170321. doi: 10.1259/dmfr.20170321. Epub 2017 Nov 6.

Abstract

OBJECTIVES

To investigate the relationship in dental cone-beam CT (CBCT) between the manufacturer-reported image pixel data and a modified conversion to CT number densities in Hounsfield unit (HU).

METHODS

A standardized CT phantom was imaged using typical clinical parameters on CBCT from three manufacturers (Carestream 9300, Carestream Health, Rochester, NY; J Morita 3D Accutomo, J. Morita Mfg. Corp., Kyoto, Japan; and Planmeca Promax 3D, Planmeca Helsinki, OY, Finland). Reconstructed axial slices were evaluated using regions of interest to ascertain the mean pixel value in five materials in the phantom. The Digital Imaging and Communications in Medicine data were also evaluated to determine if raw pixel data had been adjusted during the image reconstruction. A modified version of the existing manual HU conversion technique was applied, and the resultant slope and y-intercept were used to scale the pixel values ultimately to HU for all images.

RESULTS

The DICOM header data show that a modified rescale y-intercept was applied to both the Carestream and Planmeca image data yielding manufacturer-produced results in HU. The Morita pixel data were unmodified and report in shades of grey or grey values (GV). The Carestream manufacturer-derived HU measurements showed good correlation in air (-1000 HU), but all other materials ranged from 2.6 to 13.5 σ from the specified phantom value. Results in the modified conversion technique images fell within 1.0-2.4 σ from the specified phantom values.

CONCLUSIONS

While more studies are needed to test for regularity, this study suggests that our modified technique could be a means of getting more accurate quantitative data from dental CBCTs.

摘要

目的

研究制造商报告的图像像素数据与经改良后转换为 CT 号密度(HU)之间在口腔锥形束 CT(CBCT)中的关系。

方法

使用三种制造商(美国 Carestream 公司的 9300、Carestream Health、Rochester,NY;日本 J Morita 公司的 3D Accutomo 和芬兰 Planmeca 公司的 Promax 3D)的典型临床参数对标准化 CT 体模进行 CBCT 成像。使用感兴趣区域评估重建的轴位切片,以确定体模中五种材料的平均像素值。还评估了数字成像和通信在医学数据,以确定在图像重建过程中是否调整了原始像素数据。应用了现有的手动 HU 转换技术的改良版本,并将所得斜率和 y 截距用于对所有图像的像素值进行缩放,最终转换为 HU。

结果

DICOM 标题数据显示,对 Carestream 和 Planmeca 图像数据应用了改良的重新缩放 y 截距,从而产生 HU 中的制造商生成结果。Morita 像素数据未经过修改,以灰度或灰度值(GV)报告。Carestream 制造商衍生的 HU 测量值在空气(-1000 HU)中显示出良好的相关性,但所有其他材料的范围从指定体模值的 2.6 到 13.5 σ。改良转换技术图像的结果在指定体模值的 1.0-2.4 σ 范围内。

结论

虽然需要进行更多研究以测试规律性,但本研究表明,我们的改良技术可能是从口腔 CBCT 获得更准确定量数据的一种手段。

相似文献

1
Investigation of dental cone-beam CT pixel data and a modified method for conversion to Hounsfield unit (HU).
Dentomaxillofac Radiol. 2018 Feb;47(2):20170321. doi: 10.1259/dmfr.20170321. Epub 2017 Nov 6.
2
A simplified low-cost phantom for image quality assessment of dental cone beam computed tomography unit.
J Med Radiat Sci. 2024 Mar;71(1):78-84. doi: 10.1002/jmrs.738. Epub 2023 Nov 15.
3
Deriving Hounsfield units using grey levels in cone beam computed tomography.
Dentomaxillofac Radiol. 2010 Sep;39(6):323-35. doi: 10.1259/dmfr/19603304.
4
Image quality and absorbed dose comparison of single- and dual-source cone-beam computed tomography.
J Appl Clin Med Phys. 2018 May;19(3):360-366. doi: 10.1002/acm2.12328. Epub 2018 Apr 17.
6
Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT.
Clin Oral Implants Res. 2015;26(1):e1-7. doi: 10.1111/clr.12315. Epub 2013 Dec 11.
8
9
Automated measurement for image distortion analysis in 2D panoramic imaging of dental CBCT system: A phantom study.
Radiography (Lond). 2023 May;29(3):533-538. doi: 10.1016/j.radi.2023.02.028. Epub 2023 Mar 11.
10
Variation in voxel value distribution and effect of time between exposures in six CBCT units.
Dentomaxillofac Radiol. 2014;43(4):20130376. doi: 10.1259/dmfr.20130376. Epub 2014 Feb 3.

引用本文的文献

2
Can gray values be converted to Hounsfield units? A systematic review.
Dentomaxillofac Radiol. 2022 Jan 1;51(1):20210140. doi: 10.1259/dmfr.20210140. Epub 2021 Jun 19.

本文引用的文献

2
Identifying MRONJ-affected bone with digital fusion of functional imaging (FI) and cone-beam computed tomography (CBCT): case reports and hypothesis.
Oral Surg Oral Med Oral Pathol Oral Radiol. 2017 Mar;123(3):e106-e116. doi: 10.1016/j.oooo.2016.11.007. Epub 2016 Nov 28.
3
CBCT-based bone quality assessment: are Hounsfield units applicable?
Dentomaxillofac Radiol. 2015;44(1):20140238. doi: 10.1259/dmfr.20140238.
4
American Association of Oral and Maxillofacial Surgeons position paper on medication-related osteonecrosis of the jaw--2014 update.
J Oral Maxillofac Surg. 2014 Oct;72(10):1938-56. doi: 10.1016/j.joms.2014.04.031. Epub 2014 May 5.
5
Bone mineral density in cone beam computed tomography: Only a few shades of gray.
World J Radiol. 2014 Aug 28;6(8):607-12. doi: 10.4329/wjr.v6.i8.607.
6
Re: Management of antiresorptive osteonecrosis of the jaws with primary surgical resection.
J Oral Maxillofac Surg. 2014 Sep;72(9):1641. doi: 10.1016/j.joms.2014.05.009.
7
Scanner and kVp dependence of measured CT numbers in the ACR CT phantom.
J Appl Clin Med Phys. 2013 Nov 4;14(6):4417. doi: 10.1120/jacmp.v14i6.4417.
8
Deriving Hounsfield units using grey levels in cone beam CT: a clinical application.
Dentomaxillofac Radiol. 2012 Sep;41(6):500-8. doi: 10.1259/dmfr/31640433. Epub 2012 Jun 29.
9
Quantification of metal artifacts on cone beam computed tomography images.
Clin Oral Implants Res. 2013 Aug;24 Suppl A100:94-9. doi: 10.1111/j.1600-0501.2011.02382.x. Epub 2011 Dec 15.
10
Utility of the computed tomography indices on cone beam computed tomography images in the diagnosis of osteoporosis in women.
Imaging Sci Dent. 2011 Sep;41(3):101-6. doi: 10.5624/isd.2011.41.3.101. Epub 2011 Sep 15.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验