Environmental Health & Radiation Safety, University of Pennsylvania, Philadelphia, PA, USA.
DMD Candidate Penn Dental Medicine, University of Pennsylvania, Philadelphia, PA, USA.
Dentomaxillofac Radiol. 2018 Feb;47(2):20170321. doi: 10.1259/dmfr.20170321. Epub 2017 Nov 6.
To investigate the relationship in dental cone-beam CT (CBCT) between the manufacturer-reported image pixel data and a modified conversion to CT number densities in Hounsfield unit (HU).
A standardized CT phantom was imaged using typical clinical parameters on CBCT from three manufacturers (Carestream 9300, Carestream Health, Rochester, NY; J Morita 3D Accutomo, J. Morita Mfg. Corp., Kyoto, Japan; and Planmeca Promax 3D, Planmeca Helsinki, OY, Finland). Reconstructed axial slices were evaluated using regions of interest to ascertain the mean pixel value in five materials in the phantom. The Digital Imaging and Communications in Medicine data were also evaluated to determine if raw pixel data had been adjusted during the image reconstruction. A modified version of the existing manual HU conversion technique was applied, and the resultant slope and y-intercept were used to scale the pixel values ultimately to HU for all images.
The DICOM header data show that a modified rescale y-intercept was applied to both the Carestream and Planmeca image data yielding manufacturer-produced results in HU. The Morita pixel data were unmodified and report in shades of grey or grey values (GV). The Carestream manufacturer-derived HU measurements showed good correlation in air (-1000 HU), but all other materials ranged from 2.6 to 13.5 σ from the specified phantom value. Results in the modified conversion technique images fell within 1.0-2.4 σ from the specified phantom values.
While more studies are needed to test for regularity, this study suggests that our modified technique could be a means of getting more accurate quantitative data from dental CBCTs.
研究制造商报告的图像像素数据与经改良后转换为 CT 号密度(HU)之间在口腔锥形束 CT(CBCT)中的关系。
使用三种制造商(美国 Carestream 公司的 9300、Carestream Health、Rochester,NY;日本 J Morita 公司的 3D Accutomo 和芬兰 Planmeca 公司的 Promax 3D)的典型临床参数对标准化 CT 体模进行 CBCT 成像。使用感兴趣区域评估重建的轴位切片,以确定体模中五种材料的平均像素值。还评估了数字成像和通信在医学数据,以确定在图像重建过程中是否调整了原始像素数据。应用了现有的手动 HU 转换技术的改良版本,并将所得斜率和 y 截距用于对所有图像的像素值进行缩放,最终转换为 HU。
DICOM 标题数据显示,对 Carestream 和 Planmeca 图像数据应用了改良的重新缩放 y 截距,从而产生 HU 中的制造商生成结果。Morita 像素数据未经过修改,以灰度或灰度值(GV)报告。Carestream 制造商衍生的 HU 测量值在空气(-1000 HU)中显示出良好的相关性,但所有其他材料的范围从指定体模值的 2.6 到 13.5 σ。改良转换技术图像的结果在指定体模值的 1.0-2.4 σ 范围内。
虽然需要进行更多研究以测试规律性,但本研究表明,我们的改良技术可能是从口腔 CBCT 获得更准确定量数据的一种手段。