Bhatia Ajay K, Yabrodi Mouhammad, Carroll Mallory, Bunting Silvia, Kanter Kirk, Maher Kevin O, Deshpande Shriprasad R
Division of Pediatric Cardiology, Emory University School of Medicine, Children's Healthcare of Atlanta, Atlanta, GA 30322, United States.
Mechanical Circulatory Support Program, Children's Healthcare of Atlanta, Atlanta, GA 30322, United States.
World J Cardiol. 2017 Sep 26;9(9):749-756. doi: 10.4330/wjc.v9.i9.749.
To assess utility and correlation of known anticoagulation parameters in the management of pediatric ventricular assist device (VAD).
Retrospective study of pediatric patients supported with a Berlin EXCOR VAD at a single pediatric tertiary care center during a single year.
We demonstrated associations between activated thromboplastin time (aPTT) and R-thromboelastography (R-TEG) values (rs = 0.65, < 0.001) and between anti-Xa assay and R-TEG values (rs = 0.54, < 0.001). The strongest correlation was seen between aPTT and anti-Xa assays (rs = 0.71, < 0.001). There was also a statistically significant correlation between platelet counts and the maximum amplitude of TEG (rs = 0.71, < 0.001). Importantly, there was no association between dose of unfractionated heparin and either measure of anticoagulation (aPTT, anti-Xa or R-TEG value).
This study suggests that while there is strong correlation between aPTT, anti-Xa assay and R-TEG values for patients requiring VAD support, there is a lack of relevant correlation between heparin dose and degree of effect. This raises concern as various guidelines continue to recommend using these parameters to titrate heparin therapy.
评估已知抗凝参数在小儿心室辅助装置(VAD)管理中的效用及相关性。
对某单一儿科三级护理中心在某一年中接受柏林EXCOR VAD支持的儿科患者进行回顾性研究。
我们证明了活化部分凝血活酶时间(aPTT)与R-血栓弹力图(R-TEG)值之间存在关联(rs = 0.65,P < 0.001),抗Xa测定与R-TEG值之间也存在关联(rs = 0.54,P < 0.001)。aPTT与抗Xa测定之间的相关性最强(rs = 0.71,P < 0.001)。血小板计数与血栓弹力图的最大振幅之间也存在统计学上的显著相关性(rs = 0.71,P < 0.001)。重要的是,普通肝素剂量与任何抗凝指标(aPTT、抗Xa或R-TEG值)之间均无关联。
本研究表明,对于需要VAD支持的患者,aPTT、抗Xa测定和R-TEG值之间存在强相关性,但肝素剂量与抗凝效果之间缺乏相关联系。鉴于各种指南仍继续推荐使用这些参数来调整肝素治疗,这引发了人们的担忧。