Suppr超能文献

新型小孔巩膜植入术矫正不规则性角膜散光。

New pinhole sulcus implant for the correction of irregular corneal astigmatism.

机构信息

From the Department of Ophthalmology (C.C.Trindade, Santhiago), University of São Paulo, São Paulo, Ciencias Medicas Medical School (B.C. Trindade) and Cançado-Trindade Eye Institute (F.C.Trindade), Belo Horizonte, Brazil; the Moran Eye Center (Werner), Salt Lake City, Utah, and the University of Cincinnati (Osher), Cincinnati, Ohio, USA.

From the Department of Ophthalmology (C.C.Trindade, Santhiago), University of São Paulo, São Paulo, Ciencias Medicas Medical School (B.C. Trindade) and Cançado-Trindade Eye Institute (F.C.Trindade), Belo Horizonte, Brazil; the Moran Eye Center (Werner), Salt Lake City, Utah, and the University of Cincinnati (Osher), Cincinnati, Ohio, USA.

出版信息

J Cataract Refract Surg. 2017 Oct;43(10):1297-1306. doi: 10.1016/j.jcrs.2017.09.014.

Abstract

PURPOSE

To evaluate the effect on visual acuity of the implantation of a new intraocular pinhole device (Xtrafocus) in cases of irregular corneal astigmatism with significant visual impairment.

SETTING

University of São Paulo, São Paulo, Brazil.

DESIGN

Prospective case series.

METHODS

Pseudophakic eyes of patients with irregular corneal astigmatism were treated with the pinhole device. The causes of irregular corneal astigmatism were keratoconus, post radial keratotomy (RK), post-penetrating keratoplasty (PKP), and traumatic corneal laceration. The device was implanted in the ciliary sulcus in a piggyback configuration to minimize the effect of corneal aberrations. Preoperative and postoperative visual parameters were compared. The main outcome variables were manifest refraction, uncorrected and corrected distance and near visual acuities, subjective patient satisfaction, and intraoperative and postoperative adverse events and complications.

RESULTS

Twenty-one patients (ages 35 to 85 years) were included. There was statistically significant improvement in uncorrected and corrected (CDVA) distance visual acuities. The median CDVA improved from 20/200 (range 20/800 to 20/60) preoperatively to 20/50 (range 20/200 to 20/20) in the first month postoperatively and remained stable over the following months. Manifest refraction remained unchanged, while a subjective visual performance questionnaire revealed perception of improvement in all the tested working distances. No major complication was observed. One case presented with decentration of the device, which required an additional surgical intervention.

CONCLUSIONS

The intraocular pinhole device performed well in patients with irregular astigmatism caused by keratoconus, RK, PKP, and traumatic corneal laceration. There was marked improvement in visual function, with high patient satisfaction.

摘要

目的

评估新型眼内针孔装置(Xtrafocus)植入术对伴有明显视力损害的不规则角膜散光患者的视力的影响。

地点

巴西圣保罗大学。

设计

前瞻性病例系列研究。

方法

对伴有不规则角膜散光的白内障术后患者采用针孔装置进行治疗。导致不规则角膜散光的原因包括圆锥角膜、放射状角膜切开术(RK)后、穿透性角膜移植(PKP)后和外伤性角膜裂伤。将该装置以叠瓦式配置植入睫状沟,以最大程度地减少角膜像差的影响。比较术前和术后的视觉参数。主要观察指标为:明显屈光不正、未矫正和矫正的远、近视力、患者主观满意度以及术中、术后不良事件和并发症。

结果

共纳入 21 名患者(年龄 35 岁至 85 岁)。未矫正和矫正(CDVA)远视力均有显著改善。术后 1 个月时,未矫正的 CDVA 中位数从术前的 20/200(20/800 至 20/60)提高至 20/50(20/200 至 20/20),此后数月内保持稳定。明显屈光不正保持不变,而主观视觉表现问卷显示所有测试工作距离的视觉改善感知率均较高。未观察到重大并发症。1 例出现装置偏心,需要进一步手术干预。

结论

眼内针孔装置在圆锥角膜、RK、PKP 和外伤性角膜裂伤引起的不规则散光患者中效果良好。视觉功能明显改善,患者满意度高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验