Zboril M, Luszik-Bhadra M, Radeck D, Reginatto M, Zimbal A
Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, Braunschweig, Germany.
Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):98-101. doi: 10.1093/rpd/ncx242.
The neutron field of the PTB Thermal Neutron Calibration Facility was characterised with the help of the Bonner sphere spectrometer of PTB, NEutron MUltisphere Spectrometer (NEMUS). For the analysis of the Bonner sphere data an analytical model of the neutron energy distribution was used. The unfolding of neutron spectra was performed via the method of Bayesian parameter estimation. Comparison of the results of thermal neutron fluence and thermal ambient dose equivalent with the results of the reference method of gold activation showed a very good agreement. For fast neutrons the results obtained with NEMUS will serve as reference values for this facility. The facility is ready for use as a thermal neutron reference field for calibrations of neutron measuring devices.
在德国物理技术研究院(PTB)热中子校准设施的中子场,借助PTB的邦纳球谱仪——中子多球谱仪(NEMUS)进行了表征。为了分析邦纳球数据,使用了中子能量分布的分析模型。通过贝叶斯参数估计方法进行中子谱的解谱。热中子注量和热环境剂量当量的结果与金活化参考方法的结果比较显示出非常好的一致性。对于快中子,用NEMUS获得的结果将作为该设施的参考值。该设施已准备好用作中子测量装置校准的热中子参考场。