Muyor José M, Arrabal-Campos Francisco M, Martínez-Aparicio Carmen, Sánchez-Crespo Antonia, Villa-Pérez Manuel
Laboratory of Kinesiology, Biomechanics and Ergonomics (KIBIOMER), University of Almería, Almería, Spain.
Higher Engineering School, University of Almería, Almería, Spain.
J Back Musculoskelet Rehabil. 2017 Nov 6;30(6):1319-1325. doi: 10.3233/BMR-169750. Epub 2017 Jun 12.
To evaluate the test-retest reliability and validity of the MOCAP system for measuring spinal sagittal thoracic and lumbar curvatures and sacral inclination in a standing posture.
Twenty-five male adults were evaluated on lateral standing radiographs. The thoracic and lumbar curvatures were calculated by Harrison's posterior tangent method. The sacral inclination was defined as the angle between the tangent line of the sacral and vertical plane. In addition, MOCAP was used to calculate the spinal curvatures and sacral inclination.
The thoracic and lumbar spine and sacral inclination demonstrated excellent reliability, with mean ICCs levels greater than 0.980 and low CVs (mean: 2.15%). Systematic biases were not significant and were very near 0, and the mean standard errors were 0.257∘. ANOVA of the radiographic and MOCAP measures did not report any statistically significant differences in the comparisons. The systematic biases and mean random errors were lower than 1∘, with CVs lower than 5% and mean ICCs higher than 0.90 between sessions.
The MOCAP system delivered consistently reliable and valid results for standing curvatures compared with a radiographic technique. This system could be used with confidence in research and clinical environments for sagittal spinal curvature measurements.
评估MOCAP系统在测量站立姿势下脊柱矢状面胸段和腰段曲度以及骶骨倾斜度方面的重测信度和效度。
对25名成年男性进行站立位侧位X线片评估。胸段和腰段曲度采用哈里森后切线法计算。骶骨倾斜度定义为骶骨切线与垂直平面之间的夹角。此外,使用MOCAP计算脊柱曲度和骶骨倾斜度。
胸段和腰段脊柱以及骶骨倾斜度显示出极佳的信度,平均组内相关系数(ICC)水平大于0.980,变异系数(CV)较低(均值:2.15%)。系统偏差不显著且非常接近0,平均标准误差为0.257°。X线片测量值与MOCAP测量值的方差分析在比较中未报告任何统计学上的显著差异。系统偏差和平均随机误差低于1°,各测量期间的CV低于5%,平均ICC高于0.90。
与X线片技术相比,MOCAP系统在测量站立位曲度方面能提供一致可靠且有效的结果。该系统可在研究和临床环境中放心用于矢状面脊柱曲度测量。