通过优化脉冲序列参数减少钛腔内近距离放射治疗施源器引起的MRI信号失真。

Reduction of MRI signal distortion from titanium intracavitary brachytherapy applicator by optimizing pulse sequence parameters.

作者信息

Sullivan Thomas P, Harkenrider Matthew M, Surucu Murat, Wood Abbie M, Yacoub Joseph H, Shea Steven M

机构信息

Department of Radiology, Stritch School of Medicine, Loyola University Chicago, Maywood, IL.

Department of Radiation Oncology, Stritch School of Medicine, Loyola University Chicago, Maywood, IL.

出版信息

Brachytherapy. 2018 Mar-Apr;17(2):377-382. doi: 10.1016/j.brachy.2017.10.013. Epub 2017 Nov 22.

Abstract

PURPOSE

To demonstrate that optimized pulse sequence parameters for a T2-weighted (T2w) fast spin echo acquisition reduced artifacts from a titanium brachytherapy applicator compared to conventional sequence parameters.

METHODS AND MATERIALS

Following Institutional Review Board approval and informed consent, seven patients were successfully imaged with both standard sagittal T2w fast spin echo parameters (voxel size of 0.98 × 0.78 × 4.0 mm; readout bandwidth of 200 Hz/px; repetition time of 2800 ms; echo time of 91 ms; echo train length of 15; 36 slices; and imaging time of 3:16 min) and an additional optimized T2w sequence (voxel size of 0.98 × 0.98 × 4.0 mm; readout bandwidth of 500 Hz/px; repetition time of 3610 ms; echo time of 91 ms; echo train length of 25; 18-36 slices; and imaging time of 1:15-2:30 min), which had demonstrated artifact reduction in prior phantom work. Visualized intracavitary tandem was hand-segmented by two of the authors. Three body imaging radiologists assessed image quality and intraobserver agreement scores were analyzed.

RESULTS

The average segmented volume of the intracavitary applicator significantly (p < 0.05) decreased with the experimental pulse sequence parameters as compared to the standard pulse sequence. Comparison of experimental and standard T2w sequence qualitative scores for each reviewer showed no significant differences between the two techniques.

CONCLUSIONS

This study demonstrated that pulse sequence parameter optimization can significantly reduce distortion artifact from titanium applicators while maintaining image quality and reasonable imaging times.

摘要

目的

证明与传统序列参数相比,用于T2加权(T2w)快速自旋回波采集的优化脉冲序列参数可减少来自钛近距离放射治疗施源器的伪影。

方法和材料

经机构审查委员会批准并获得知情同意后,对7名患者成功进行了标准矢状面T2w快速自旋回波参数(体素大小为0.98×0.78×4.0 mm;读出带宽为200 Hz/像素;重复时间为2800 ms;回波时间为91 ms;回波链长度为15;36层;成像时间为3:16分钟)以及另外一种优化的T2w序列(体素大小为0.98×0.98×4.0 mm;读出带宽为500 Hz/像素;重复时间为3610 ms;回波时间为91 ms;回波链长度为25;18 - 36层;成像时间为1:15 - 2:30分钟)的成像,该优化序列在之前的体模研究中已证明可减少伪影。两名作者对可视化的腔内串联装置进行手动分割。三名体部成像放射科医生评估图像质量,并分析观察者内一致性评分。

结果

与标准脉冲序列相比,采用实验性脉冲序列参数时,腔内施源器的平均分割体积显著减小(p < 0.05)。每位审阅者对实验性和标准T2w序列定性评分的比较显示,两种技术之间无显著差异。

结论

本研究表明,脉冲序列参数优化可显著减少钛制施源器产生的畸变伪影,同时保持图像质量和合理的成像时间。