Suppr超能文献

[鼻部和耳部传统体层摄影术与高分辨率计算机断层扫描术的辐射剂量及临床价值]

[Radiation dosage and clinical value of conventional tomography and high resolution computerized tomography of the nose and ear].

作者信息

Grobovschek M, Rahim H, Oberascher G

机构信息

Röntgenabteilung, Landesnervenklinik Salzburg.

出版信息

HNO. 1989 Jan;37(1):23-6.

PMID:2917879
Abstract

Hypocycloidal (frontal and temporal) tomography of the base of the skull is compared with high resolution computed tomography (HR-CT) with respect to dose and clinical usefulness. In most cases HR-CT is indicated because of its better density resolution in comparison with conventional hypocycloidal tomography. As the X-ray beam is directed at the slice plane the dosage is noticeably less than when using conventional tomography. Slice thickness depends on the required geometric resolution. This is an additional factor in radiation protection.

摘要

对头骨底部的摆线(额部和颞部)体层摄影术与高分辨率计算机断层扫描(HR-CT)在剂量和临床实用性方面进行了比较。在大多数情况下,由于与传统摆线体层摄影术相比,HR-CT具有更好的密度分辨率,因此建议使用HR-CT。由于X射线束直接指向切片平面,其剂量明显低于使用传统体层摄影术时的剂量。切片厚度取决于所需的几何分辨率。这是辐射防护中的一个额外因素。