Clinical Biochemistry section, Department of Neurological, Biomedical and Movement Sciences, University of Verona, Italy.
DASP s.r.l., Gerenzano, Italy.
Biochem Med (Zagreb). 2017 Oct 15;27(3):031001. doi: 10.11613/BM.2017.031001.
Liquid chromatography coupled to atmospheric pressure ionization tandem mass spectrometry (LC-ESI-MS/MS) is currently considered the reference method for quantitative determination of urinary free cortisol (UFC). One of the major drawbacks of this measurement is a particular form of matrix effect, conventionally known as ion suppression.
We describe here the case of a 66-year-old-patient referred to the daily service of general medicine for intravenous antibiotic administration due to a generalized infection and for routine 24 hours UFC monitoring in the setting of glucocorticoid replacement therapy.
The observation of 10-fold decrease of internal standard of cortisol signal led us to hypothesize the presence of an ion suppression effect due to a co-eluting endogenous compound. Screening analysis of tandem mass spectrometry (MS/MS) spectra of the interfering molecule, along with confirmation analyses, were suggestive of the presence of high concentration of piperacillin. The problem was then easily solved with minor modifications of the chromatographic technique.
According to our findings, antibiotic therapy with piperacillin/tazobactam should be regarded as an important interference in UFC assessment, which may potentially affect detection capability, precision and accuracy of this measurement. This case report emphasizes that accurate anamnesis and standardization of all phases of urine collection are essential aspects for preventing potential interference in laboratory testing.
液相色谱-大气压电离串联质谱(LC-ESI-MS/MS)目前被认为是定量测定尿游离皮质醇(UFC)的参考方法。这种测量方法的一个主要缺点是一种特殊形式的基质效应,通常称为离子抑制。
我们在这里描述了一位 66 岁患者的情况,该患者因全身感染而被转至普通内科的日常门诊,接受静脉注射抗生素治疗,并在糖皮质激素替代治疗的情况下进行常规 24 小时 UFC 监测。
观察到皮质醇内标信号降低了 10 倍,这使我们假设存在由于共洗脱内源性化合物而导致的离子抑制效应。对内源性干扰物质的串联质谱(MS/MS)谱进行筛选分析,并进行确认分析,提示存在高浓度哌拉西林。通过对色谱技术进行微小修改,问题很容易得到解决。
根据我们的发现,哌拉西林/他唑巴坦的抗生素治疗应被视为 UFC 评估中的一个重要干扰因素,这可能会潜在地影响该测量的检测能力、精密度和准确性。本病例报告强调,准确的病史和尿液采集所有阶段的标准化是防止实验室检测中潜在干扰的重要方面。