Dunbar M J, Laende E K, Collopy D, Richardson C G
Dalhousie University, Division of Orthopaedics, Department of Surgery, Dalhousie University, and Halifax Infirmary, QEII Health Sciences Centre, Nova Scotia Health Authority, 1796 Summer St, Halifax NS B3H 3A7, Canada.
Halifax Infirmary, Division of Orthopaedics, Department of Surgery, Dalhousie University, and Halifax Infirmary, QEII Health Sciences Centre, Nova Scotia Health Authority, 1796 Summer St., Halifax NS B3H 3A7, Canada.
Bone Joint J. 2017 Dec;99-B(12):1596-1602. doi: 10.1302/0301-620X.99B12.BJJ-2016-1118.R2.
Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation.
A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component.
While there was significant variation in individual migration patterns, the overall migration of the tibial component in the study group demonstrated a pattern of initial migration followed by stabilisation after one year, with mean maximum total point motion (MTPM) of 0.02 mm (standard deviation (sd) 0.20) between one and two years post-operatively. The direction of greatest motion was subsidence, which stabilised at three months post-operatively (mean translation of 0.21 mm, sd 0.40).
The tibial component migration pattern of stabilisation in the second post-operative year is indicative of successful long-term fixation for this PA-coated tibial component. Cite this article: 2017;99-B:1596-1602.
用于全膝关节置换术中非骨水泥固定的羟基磷灰石涂层理论上可与宿主骨形成持久的生物界面。本研究的目的是使用放射性立体测量分析测量的部件迁移作为长期固定的指标,用合适的羟基磷灰石、磷灰石涂层胫骨部件来验证这一假设。
两个中心共有29例患者接受了非骨水泥磷灰石涂层胫骨部件,并通过放射性立体测量分析检查随访两年,以量化部件的迁移情况。
虽然个体迁移模式存在显著差异,但研究组胫骨部件的总体迁移呈现出术后第一年先迁移然后稳定的模式,术后1至2年的平均最大总点运动(MTPM)为0.02毫米(标准差(sd)0.20)。最大运动方向为下沉,术后3个月稳定(平均平移0.21毫米,sd 0.40)。
术后第二年胫骨部件迁移稳定的模式表明该磷灰石涂层胫骨部件长期固定成功。引用本文:2017;99-B:1596-1602。