使用新型带细输送系统的金属支架进行内镜超声引导下胆道引流:在体模和猪模型中的临床前研究

Endoscopic ultrasound-guided biliary drainage using a newly designed metal stent with a thin delivery system: a preclinical study in phantom and porcine models.

作者信息

Minaga Kosuke, Kitano Masayuki, Itonaga Masahiro, Imai Hajime, Miyata Takeshi, Yamao Kentaro, Tamura Takashi, Nuta Junya, Warigaya Kenji, Kudo Masatoshi

机构信息

Department of Gastroenterology and Hepatology, Kindai University Faculty of Medicine, Osaka-Sayama, Japan.

Second Department of Internal Medicine, Wakayama Medical University, 811-1 Kimiidera, Wakayama, 641-8509, Japan.

出版信息

J Med Ultrason (2001). 2018 Jul;45(3):391-397. doi: 10.1007/s10396-017-0850-1. Epub 2017 Dec 8.

Abstract

PURPOSE

This study was designed to evaluate the feasibility and safety of a newly designed self-expandable metal stent for endoscopic ultrasound-guided biliary drainage (EUS-BD) when it was delivered via three different stent delivery systems: a 7.5Fr delivery catheter with a bullet-shaped tip (7.5Fr-bullet), a 7Fr catheter with a bullet-shaped tip (7Fr-bullet), or a 7Fr catheter with a tee-shaped tip (7Fr-tee).

METHODS

This experimental study utilized a porcine model of biliary dilatation involving ten pigs. In the animal study, technical feasibility and clinical outcomes of the stent when placed with each of the delivery systems were examined. In addition, a phantom model was used to measure the resistance of these delivery systems to advancement.

RESULTS

Phantom experiments showed that, compared with 7Fr-bullet, 7Fr-tee had less resistance force to the advancement of the stent delivery system. EUS-BD was technically successful in all ten pigs. Fistulous tract dilation was necessary in 100% (2/2), 75% (3/4), and 0% (0/4) of the pigs that underwent EUS-BD using 7.5Fr-bullet, 7Fr-bullet, and 7Fr-tee, respectively. There were no procedure-related complications.

CONCLUSION

Our newly designed metal stent may be feasible and safe for EUS-BD, particularly when delivered by 7Fr-tee, because it eliminates the need for fistulous tract dilation.

摘要

目的

本研究旨在评估一种新设计的自膨式金属支架用于内镜超声引导下胆道引流(EUS-BD)时,通过三种不同支架输送系统输送的可行性和安全性:一种带有子弹头形状尖端的7.5Fr输送导管(7.5Fr-子弹头)、一种带有子弹头形状尖端的7Fr导管(7Fr-子弹头)或一种带有T形尖端的7Fr导管(7Fr-T形)。

方法

本实验研究采用了涉及10头猪的胆道扩张猪模型。在动物研究中,检查了使用每种输送系统放置支架时的技术可行性和临床结果。此外,使用了一个模型来测量这些输送系统推进时的阻力。

结果

模型实验表明,与7Fr-子弹头相比,7Fr-T形对支架输送系统推进的阻力较小。EUS-BD在所有10头猪中技术上均获成功。分别使用7.5Fr-子弹头、7Fr-子弹头和7Fr-T形进行EUS-BD的猪中,100%(2/2)、75%(3/4)和0%(0/4)需要进行瘘道扩张。没有与操作相关的并发症。

结论

我们新设计的金属支架用于EUS-BD可能是可行且安全的,特别是通过7Fr-T形输送时,因为它无需进行瘘道扩张。