Verma Teerth Raj, Painuly Nirmal K, Mishra Surendra P, Singh Navin, Bhatt M L B, Jamal Naseem, Pant M C
Department of Radiotherapy, King George Medical University UP, Lucknow, Uttart Pradesh, India.
Department of Radiotherapy, Dr. Ram Manohar Lohia Institute of Medical Sciences, Lucknow, Uttart Pradesh, India.
J Cancer Res Ther. 2017 Oct-Dec;13(6):1007-1014. doi: 10.4103/0973-1482.168992.
To evaluate dose calculation accuracy of various algorithms in lung equivalent inhomogeneity comprising tumor within it and comparison with Gafchromic film data.
Gafchromic film measured central axis absorbed dose in lung insert (-700 Hounsfield unit [HU]), in racemosa wood cylindrical inhomogeneity (-725 HU) and at three surfaces of tumor (-20 HU) created in cylindrical inhomogeneity, put in the cavity of computerized imaging reference systems (CIRS) thorax phantom were compared with convolution (CON), superposition (SP), fast SP (FSP), and X-ray voxel Monte Carlo (XVMC) algorithms calculated dose using 6 MV beams of field size 2 cm × 2 cm, 3 cm × 3 cm, 4 cm × 4 cm, 5 cm × 5 cm, and 8 cm × 8 cm.
XVMC was in good agreement with film measured results for all selected field sizes except 3 cm × 3 cm. SP under estimated by 5.7% at the center of the lung insert while deviation up to 6% was found at the cent of wood inhomogeneity in 2 cm × 2 cm. Except CON, increase in dose from proximal to the central surface of the tumor and then dose falloff from central to the distal surface for field size 2 cm × 2 cm to 4 cm × 4 cm was recorded. The change in film measured percentage depth dose from 2 cm × 2 cm to 3 cm × 3 cm field sizes was found -8% however for consecutive field size(s) larger than 3 cm × 3 cm this difference was less. CON and FSP produced overestimated results.
Out of four algorithms, XVMC found consistent with measured data. The electronic disequilibrium within and at the interface of inhomogeneity make the accurate dose predictions difficult. These limitations results in deviations from the expected results of the treatments.
评估各种算法在包含肿瘤的肺等效不均匀性中的剂量计算准确性,并与放射变色胶片数据进行比较。
将置于计算机成像参考系统(CIRS)胸部体模腔内的放射变色胶片测量的肺插入物(-700亨氏单位[HU])、总状土木圆柱形不均匀性(-725 HU)以及在圆柱形不均匀性中创建的肿瘤三个表面(-20 HU)的中心轴吸收剂量,与卷积(CON)、叠加(SP)、快速叠加(FSP)和X射线体素蒙特卡罗(XVMC)算法使用6 MV射束、射野尺寸为2 cm×2 cm、3 cm×3 cm、4 cm×4 cm、5 cm×5 cm和8 cm×8 cm计算的剂量进行比较。
除3 cm×3 cm射野外,XVMC与胶片测量结果在所有选定射野尺寸下均具有良好的一致性。SP在肺插入物中心低估了5.7%,而在2 cm×2 cm的木材不均匀性中心发现偏差高达6%。除CON外,记录了射野尺寸从2 cm×2 cm到4 cm×4 cm时,从肿瘤近端到中心表面剂量增加,然后从中心到远端表面剂量下降的情况。发现从2 cm×2 cm到3 cm×3 cm射野尺寸时,胶片测量的百分深度剂量变化为-8%,然而对于大于3 cm×3 cm的连续射野尺寸,这种差异较小。CON和FSP产生了高估结果。
在四种算法中,XVMC与测量数据一致。不均匀性内部和界面处的电子不平衡使得准确的剂量预测变得困难。这些限制导致与预期治疗结果存在偏差。