立体定向、单次剂量肺部肿瘤放射治疗:基于实际患者 CT 扫描的笔形束和蒙特卡罗算法之间的绝对剂量和剂量分布比较。
Stereotactic, single-dose irradiation of lung tumors: a comparison of absolute dose and dose distribution between pencil beam and Monte Carlo algorithms based on actual patient CT scans.
机构信息
Department of Radiation Oncology, Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.
出版信息
Int J Radiat Oncol Biol Phys. 2010 Nov 1;78(3):955-63. doi: 10.1016/j.ijrobp.2009.08.012. Epub 2010 Feb 18.
PURPOSE
Dose calculation based on pencil beam (PB) algorithms has its shortcomings predicting dose in tissue heterogeneities. The aim of this study was to compare dose distributions of clinically applied non-intensity-modulated radiotherapy 15-MV plans for stereotactic body radiotherapy between voxel Monte Carlo (XVMC) calculation and PB calculation for lung lesions.
METHODS AND MATERIALS
To validate XVMC, one treatment plan was verified in an inhomogeneous thorax phantom with EDR2 film (Eastman Kodak, Rochester, NY). Both measured and calculated (PB and XVMC) dose distributions were compared regarding profiles and isodoses. Then, 35 lung plans originally created for clinical treatment by PB calculation with the Eclipse planning system (Varian Medical Systems, Palo Alto, CA) were recalculated by XVMC (investigational implementation in PrecisePLAN [Elekta AB, Stockholm, Sweden]). Clinically relevant dose-volume parameters for target and lung tissue were compared and analyzed statistically.
RESULTS
The XVMC calculation agreed well with film measurements (<1% difference in lateral profile), whereas the deviation between PB calculation and film measurements was up to +15%. On analysis of 35 clinical cases, the mean dose, minimal dose and coverage dose value for 95% volume of gross tumor volume were 1.14 ± 1.72 Gy, 1.68 ± 1.47 Gy, and 1.24 ± 1.04 Gy lower by XVMC compared with PB, respectively (prescription dose, 30 Gy). The volume covered by the 9 Gy isodose of lung was 2.73% ± 3.12% higher when calculated by XVMC compared with PB. The largest differences were observed for small lesions circumferentially encompassed by lung tissue.
CONCLUSIONS
Pencil beam dose calculation overestimates dose to the tumor and underestimates lung volumes exposed to a given dose consistently for 15-MV photons. The degree of difference between XVMC and PB is tumor size and location dependent. Therefore XVMC calculation is helpful to further optimize treatment planning.
目的
基于笔束(PB)算法的剂量计算在预测组织异质性中的剂量方面存在不足。本研究旨在比较临床应用的 15-MV 非调强放疗立体定向体部放疗计划在肺病变中,体素蒙特卡罗(XVMC)计算和 PB 计算的剂量分布。
方法与材料
为了验证 XVMC,我们在一个不均匀的胸部体模中使用 EDR2 胶片(Eastman Kodak,Rochester,NY)对一个治疗计划进行了验证。比较了测量值和计算值(PB 和 XVMC)的剂量分布,包括轮廓和等剂量曲线。然后,我们使用 Eclipse 计划系统(Varian Medical Systems,Palo Alto,CA)对 35 个最初由 PB 计算创建的临床治疗计划进行了重新计算,采用 XVMC(PrecisePLAN [Elekta AB,Stockholm,Sweden]中的研究性实现)。比较并统计分析了靶区和肺组织的临床相关剂量-体积参数。
结果
XVMC 计算与胶片测量值吻合良好(侧位轮廓差异<1%),而 PB 计算与胶片测量值的偏差高达+15%。在对 35 例临床病例的分析中,与 PB 相比,XVMC 计算得到的大体肿瘤体积 95%体积的平均剂量、最小剂量和覆盖剂量值分别低 1.14±1.72Gy、1.68±1.47Gy 和 1.24±1.04Gy(处方剂量为 30Gy)。XVMC 计算的 9Gy 等剂量线覆盖的肺体积比 PB 计算的高 2.73%±3.12%。差异最大的是肺组织周围小病变。
结论
对于 15-MV 光子,PB 剂量计算高估了肿瘤的剂量,低估了暴露于给定剂量的肺体积。XVMC 和 PB 之间的差异程度取决于肿瘤的大小和位置。因此,XVMC 计算有助于进一步优化治疗计划。