Rinard George A, Quine Richard W, McPeak Joseph, Buchanan Laura, Eaton Sandra S, Eaton Gareth R
School of Engineering and Computer Science and Center for EPR Imaging In Vivo Physiology, University of Denver, Denver, CO 80208.
Department of Chemistry and Biochemistry and Center for EPR Imaging In Vivo Physiology, University of Denver, Denver, CO 80208.
Appl Magn Reson. 2017 Dec;48(11-12):1219-1226. doi: 10.1007/s00723-017-0945-2. Epub 2017 Sep 14.
A copper X-band (9.22 GHz) cross loop resonator has been constructed for use with 4 mm sample tubes. The Q for the two resonators are 380 and 350, respectively. The resonator efficiency is about 1 G per square root of watt. Operation has been demonstrated with measurement of T by saturation recovery for samples of coal and an immobilized nitroxide radical.
已构建了一个用于4毫米样品管的铜制X波段(9.22吉赫兹)交叉环形谐振器。两个谐振器的品质因数分别为380和350。谐振器效率约为每瓦特平方根1吉。通过对煤样品和固定化氮氧化物自由基进行饱和恢复测量T,已证明了该操作。