The influence of film-screen color sensitivity and type of measurement device on kVp measurements.

作者信息

Lam R W, Price S C

机构信息

Department of Radiologic Technologies, Medical College of Georgia, Augusta 30912-0600.

出版信息

Radiol Technol. 1989 Mar-Apr;60(4):319-21.

PMID:2928498
Abstract

Three methods for evaluating radiographic kVp were studied: the Wisconsin Test Cassette, the Noninvasive Evaluator of Radiation Outputs (NERO), and the Dynalyzer. The Dynalyzer kVp readings were the highest and were followed by NERO and cassette readings in descending order. By film type, the cassette readings ranged from Kodak OG (green sensitive), TMG (green sensitive), XK (blue sensitive), and XRP (blue sensitive) in descending order. The results show that there is significant variation between the methods.

摘要

研究了三种评估射线照相千伏峰值(kVp)的方法:威斯康星测试暗盒、辐射输出无创评估仪(NERO)和剂量分析仪。剂量分析仪的kVp读数最高,其次是NERO,暗盒读数按降序排列。按胶片类型,暗盒读数从柯达OG(绿色敏感)、TMG(绿色敏感)、XK(蓝色敏感)和XRP(蓝色敏感)依次递减。结果表明,这些方法之间存在显著差异。