Suppr超能文献

高频震荡通气时吸入一氧化氮传递的台架研究。

A bench study of inhaled nitric oxide delivery during high frequency percussive ventilation.

机构信息

University of Cincinnati College of Medicine, Cincinnati, Ohio.

Mallinckrodt Pharmaceuticals, Hampton, New Jersey.

出版信息

Pediatr Pulmonol. 2018 Mar;53(3):337-341. doi: 10.1002/ppul.23934. Epub 2018 Jan 4.

Abstract

BACKGROUND

Safe and effective delivery of inhaled nitric oxide (INO) requires the appropriate interface of ventilator and INO delivery device.

METHODS

We compared INO delivery using four configurations with the Transport Sinusoidal Bronchotron® and INOmax DSIR Plus in a lung model. Ventilator settings and lung model values were held constant. Delivered NO, NO , and inspired oxygen (FIO ) were measured. The mean difference between set and measured NO was calculated and compared using ANOVA.

RESULTS

Placement of the injector module in line with the sliding venturi resulted in a ventilator failure. With both continuous flow techniques there was no appreciable NO generated and the mean difference between set NO and measured NO at 20 and 40 ppm was -16.5 ppm and -33.2 ppm at flows of 5 and 10 L/min. Placement of the injector module between the sliding venturi and lung model resulted in an increase of NO2 to a peak of 2.4 ppm (mean 2.3 + 0.1) and a mean difference between set and measured NO of + 11.3 ppm and +30 ppm at 20 and 40 ppm, 300 cycles per minute (cpm), and 22.1 ppm and 37.6 ppm, at 20 and 40 ppm, 600 cpm. None of the test configurations delivered INO within 30% of set concentrations. No alarms or interruption of INO delivery occurred.

CONCLUSION

The dual gas delivery system of the Bronchotron prevents accurate delivery of INO. The combination of these two devices should be accomplished with caution and vigilance.

摘要

背景

安全有效地输送吸入一氧化氮(INO)需要呼吸机和 INO 输送设备的适当接口。

方法

我们在肺模型中比较了四种配置使用 Transport Sinusoidal Bronchotron® 和 INOmax DSIR Plus 输送 INO 的情况。保持呼吸机设置和肺模型值不变。测量输送的 NO、NO 和吸入氧气(FIO )。使用 ANOVA 计算并比较设定和测量的 NO 之间的平均差异。

结果

将注射器模块放置在滑动文丘里管中会导致呼吸机故障。使用两种连续流动技术都没有产生可观的 NO,并且在 5 和 10 L/min 的流速下,设定的 NO 和测量的 NO 之间的平均差异为-16.5 ppm 和-33.2 ppm。将注射器模块放置在滑动文丘里管和肺模型之间会导致 NO2 增加到 2.4 ppm(平均 2.3+0.1)的峰值,并且在 300 次/分钟(cpm)时,设定和测量的 NO 之间的平均差异为+11.3 ppm 和+30 ppm,在 20 和 40 ppm 时,设定和测量的 NO 之间的平均差异为+22.1 ppm 和+37.6 ppm。在 600 cpm 时,所有测试配置均无法在 30%的设定浓度内输送 INO。没有发生 INO 输送的报警或中断。

结论

Bronchotron 的双气体输送系统会阻止 INO 的准确输送。这两种设备的组合应谨慎并保持警惕。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验