运动血氧测定在动脉狭窄性跛行中的测试内和测试间可靠性。

Intra-test and test-retest reliability of exercise oximetry in arterial claudication.

机构信息

Laboratoire d'explorations fonctionnelles vasculaires, CHU d'Angers, France.

Laboratoire d'explorations fonctionnelles vasculaires, CHU d'Angers, France; Laboratoire de Physiologie, Institut Supérieur des Sciences de la Santé, Université Nazi Boni, Bobo-Dioulasso, Burkina Faso.

出版信息

Microvasc Res. 2018 May;117:44-49. doi: 10.1016/j.mvr.2018.01.003. Epub 2018 Jan 9.

Abstract

BACKGROUND

Transcutaneous oxygen pressure (tcpO2) reliability is blunted by an unpredictable transcutaneous gradient through the skin. We hypothesized that the "Decrease from Rest of Oxygen pressure (DROP: subtraction of limb-changes from chest-changes from the respective starting values) would show a good to excellent reliability during Exercise -tcpO2 investigations.

METHODS

In three different experiments we tested: The intra-test variability at the peripheral level (Experiment A: 32 patients, 16 at each location), at the chest level (Experiment B: 45 patients) and the test-retest reproducibility within 3 months (Experiment C: 67 patients). We calculated the intra-class coefficient of correlation (ICC) with 95% confidence interval [Lower/upper limit]. ICC between 0.60 and 0.749 indicate a good agreement. ICC above 0.750 indicates an excellent agreement.

RESULTS

ICC values for DROP-min were 0.848 [0.723/0.935] at the buttock and 0.920 [0.846/0.967] at the calf levels, in experiment A; ICC were 0.873 [0.799/0.923] at the buttock and 0.898 [0.790/0.953] at the calf levels, in experiment B; 0.806 [0.716/0.871] at then buttock level (n = 67) and 0.807 [0.722/0.868] at the calf in experiment C.

CONCLUSIONS

Intra-test and test-retest reliability is excellent using the DROP calculation for exercise-tcpO2 investigations.

摘要

背景

经皮氧分压(tcpO2)的可靠性会受到皮肤中不可预测的经皮梯度的影响而减弱。我们假设“从休息时的氧压下降(DROP:从各自的起始值减去肢体变化与胸部变化)在运动-tcpO2 研究中会显示出良好到极好的可靠性。

方法

我们在三个不同的实验中进行了测试:在末梢水平的组内测试变异性(实验 A:32 名患者,每个部位 16 名),在胸部水平(实验 B:45 名患者)以及 3 个月内的测试-再测试可重复性(实验 C:67 名患者)。我们用 95%置信区间[下限/上限]计算了组内相关系数(ICC)。ICC 值在 0.60 到 0.749 之间表示良好的一致性。ICC 值大于 0.750 表示极好的一致性。

结果

在实验 A 中,DROP-min 的 ICC 值在臀部为 0.848[0.723/0.935],在小腿为 0.920[0.846/0.967];在实验 B 中,臀部为 0.873[0.799/0.923],小腿为 0.898[0.790/0.953];在实验 C 中,臀部为 0.806[0.716/0.871](n=67),小腿为 0.807[0.722/0.868]。

结论

使用 DROP 计算在运动-tcpO2 研究中具有极好的组内和测试-再测试可靠性。