Son Jaeman, Lee Se Byeong, Lim Youngkyung, Park Sung Yong, Cho Kwanho, Yoon Myonggeun, Shin Dongho
Department of Bio-convergence Engineering, Korea University, Seoul 02841, Korea.
Proton Therapy Center, National Cancer Center, Goyang 10408, Korea.
Sensors (Basel). 2018 Jan 15;18(1):227. doi: 10.3390/s18010227.
This study describes the development of a beam monitoring system for the verification of entrance dose map in pencil beam scanning (PBS) proton therapy based on fiber optic radiation sensors (FORS) and the validation of this system through a feasibility study. The beam monitoring system consisted of 128 optical fibers optically coupled to photo-multiplier tubes. The performance of the beam monitoring system based on FORS was verified by comparing 2D dose maps of square-shaped fields of various sizes, which were obtained using conventional dosimeters such as MatriXX and EBT3 film, with those measured using FORS. The resulting full-width at half maximum and penumbra were compared for PBS proton beams, with a ≤2% difference between each value, indicating that measurements using the conventional dosimetric tool corresponded to measurements based on FORS. For irregularly-shaped fields, a comparison based on the gamma index between 2D dose maps obtained using MatriXX and EBT3 film and the 2D dose map measured by the FORS showed passing rates of 96.9 ± 1.3% and 96.2 ± 1.9%, respectively, confirming that FORS-based measurements for PBS proton therapy agreed well with those measured using the conventional dosimetric tools. These results demonstrate that the developed beam monitoring system based on FORS is good candidate for monitoring the entrance dose map in PBS proton therapy.
本研究描述了一种基于光纤辐射传感器(FORS)的用于铅笔束扫描(PBS)质子治疗中入射剂量图验证的束流监测系统的开发,并通过可行性研究对该系统进行了验证。束流监测系统由128根与光电倍增管光学耦合的光纤组成。通过比较使用MatriXX和EBT3胶片等传统剂量仪获得的各种尺寸方形射野的二维剂量图与使用FORS测量的剂量图,验证了基于FORS的束流监测系统的性能。比较了PBS质子束的半高宽和半值层,每个值之间的差异≤2%,表明使用传统剂量学工具的测量结果与基于FORS的测量结果一致。对于不规则形状的射野,基于MatriXX和EBT3胶片获得的二维剂量图与FORS测量的二维剂量图之间的伽马指数比较显示通过率分别为96.9±1.3%和96.2±1.9%,证实了PBS质子治疗中基于FORS的测量结果与使用传统剂量学工具的测量结果吻合良好。这些结果表明,所开发的基于FORS的束流监测系统是监测PBS质子治疗中入射剂量图的良好候选方案。