两种直接金属激光烧结制造的牙种植体的准静态强度和断口分析。

Quasi-static strength and fractography analysis of two dental implants manufactured by direct metal laser sintering.

机构信息

Department of Research, Biotecnos Research Center, Montevideo, Uruguay.

Department of Biotechnology of the University Catholica San Antonio de Murcia (UCAM), Murcia, Spain.

出版信息

Clin Implant Dent Relat Res. 2018 Jun;20(3):368-374. doi: 10.1111/cid.12590. Epub 2018 Jan 30.

Abstract

BACKGROUND

New manufacturing methods was developed to improve the tissues integration with the titanium alloy pieces.

OBJECTIVE

The present in vitro study was to assess the resistance and fracture mode after applied a quasi-static compressive force on the two dental implants manufactured by direct metal laser sintering.

MATERIALS AND METHODS

Twenty dental implants manufactured by direct metal laser sintering, using titanium alloy (Ti-6Al-4V) granules in two designs (n = 10 per group): Conventional dental implant (group Imp1) two-piece implant design, where the surgical implant and prosthetic abutment are two separate components and, the one-piece implant (group Imp2), where the surgical implant and prosthetic abutment are one integral piece. All samples were subjected to quasi-static loading at a 30° angle to the implant axis in a universal testing machine.

RESULTS

The mean fracture strengths were 1269.2 ± 128.8 N for the group Imp1 and, 1259.5 ± 115.1 N for the group Imp2, without statistical differences (P = .8722). In both groups, the fracture surface does not present crack between the compact core and the superficial (less dense and porous) part of the implants.

CONCLUSIONS

Based on the measured resistance data for the two implant models manufactured by direct metal laser sintering tested in the present study, we can suggest that they have adequate capacity to withstand the masticatory loads.

摘要

背景

新的制造方法旨在提高钛合金部件与组织的整合。

目的

本体外研究旨在评估两种通过直接金属激光烧结制造的牙科植入物在施加准静态压缩力后的阻力和断裂模式。

材料和方法

使用钛合金(Ti-6Al-4V)颗粒,设计了两种牙科植入物(n = 每组 10 个):传统牙科植入物(组 Imp1),为两件式植入物设计,其中手术植入物和修复体基台是两个单独的部件,以及一体式植入物(组 Imp2),其中手术植入物和修复体基台是一个整体部件。所有样本均在万能试验机上以 30°角于植入物轴进行准静态加载。

结果

组 Imp1 的平均断裂强度为 1269.2 ± 128.8 N,组 Imp2 的平均断裂强度为 1259.5 ± 115.1 N,无统计学差异(P = .8722)。在两组中,断裂面均未在致密核心和植入物表面(密度较低且多孔)之间出现裂纹。

结论

基于本研究中通过直接金属激光烧结制造的两种植入物模型的测量阻力数据,我们可以建议它们具有足够的能力来承受咀嚼负荷。