Saleh Tawfik A
Chemistry Department, King Fahd University of Petroleum and Minerals, P.O. Box 6724, Dhahran 31261, Saudi Arabia.
J Pharm Anal. 2011 Nov;1(4):246-250. doi: 10.1016/j.jpha.2011.09.002. Epub 2011 Oct 24.
This paper reports on development of a system consisting of a portable sequential injector coupled with potentiometric unit for sensing of chlorpheniramine (CPA), based on the reaction of CPA with potassium permanganate in acidic media. Various experimental conditions affecting the potential intensity were studied and incorporated into the procedure. Under the optimum conditions, linear relationship between the CPA concentration and peak area was obtained for the concentration range of 0.1-50 ppm. The method reflects good recovery with relative standard deviation (RSD)<3%. The detection limit was 0.05 ppm. The developed method was successfully applied for determination of CPA in pure form and in pharmaceutical dosage forms. The results, obtained using the method, are in accord with the results of the British pharmacopoeia method. In addition to its accuracy and precision, the method has the advantages of being simple, inexpensive and rapid.
本文报道了一种基于氯苯那敏(CPA)在酸性介质中与高锰酸钾反应,由便携式顺序注射仪与电位测定单元组成的用于检测CPA的系统的开发。研究了影响电位强度的各种实验条件,并将其纳入检测程序。在最佳条件下,CPA浓度在0.1 - 50 ppm范围内与峰面积呈线性关系。该方法回收率良好,相对标准偏差(RSD)<3%。检测限为0.05 ppm。所开发的方法成功应用于纯品及药物剂型中CPA的测定。用该方法得到的结果与英国药典方法的结果一致。该方法除了具有准确性和精密度外,还具有简单、廉价和快速的优点。