Poots K, Cobbold R S, Johnston K W, Appugliese R, Kassam M, Zuech P E, Hummel R L
Ann Biomed Eng. 1986;14(3):203-18. doi: 10.1007/BF02584271.
A nonintrusive method for the visualization of pulsatile flow velocity profiles is described. The method is based on the use of a photochromic dye that is added to the fluid being studied and a nitrogen laser which excites the dye producing a marker "line" whose movement can be photographed. A microcomputer is used as a system controller, to coordinate the system timing and to manage the data transfer. The method used for analysis of the photographs to determine the velocity profiles is described. Examples are presented of instantaneous velocity profiles obtained from velocity waveforms that are similar to those of the femoral artery. In addition, application of the system for studying the relationship between Doppler ultrasound spectral recordings and flow velocity profiles is discussed.
本文描述了一种用于可视化脉动血流速度剖面的非侵入性方法。该方法基于使用一种添加到所研究流体中的光致变色染料,以及一台激发该染料产生可拍摄其移动的标记“线”的氮激光器。使用一台微型计算机作为系统控制器,以协调系统定时并管理数据传输。描述了用于分析照片以确定速度剖面的方法。给出了从类似于股动脉速度波形获得的瞬时速度剖面的示例。此外,还讨论了该系统在研究多普勒超声频谱记录与血流速度剖面之间关系方面的应用。