The BD FACSVia™ system is a novel flow cytometer with improved workflow efficiencies. To evaluate the HLA-B27 application developed on the BD FACSVia system utilizing the BD™ HLA-B27 kit, we conducted a concordance study at three centers to compare with the BD FACSCalibur™ system. Prepared donor samples (n = 594) were analyzed on both the BD FACSVia and BD FACSCalibur for the HLA-B27 assay. Adjudication of HLA-B27 discordant results was performed using the reverse sequence-specific oligonucleotide (rSSO) DNA typing method (LABType® SSO, One Lambda). On the BD FACSVia system 80 B27 positive, 499 B27 negative and 15 "Inconclusive" samples were observed. The corresponding BD FACSCalibur results were 73 B27 positive, 502 B27 negative and 19 "gray zone" samples. The overall concordance of HLA-B27 determination was 98% between the two systems with seven more positives identified on BD FACSVia as compared to BD FACSCalibur. The equivocal zone between positive and negative on BD FACSVia (named "Inconclusive") and on BD FACSCalibur (named "gray zone") is due to antibody cross reactivity of HLA-B27 clone GS145.2. One negative sample verified with the rSSO DNA method was reported as HLA-B27 positive by the BD FACSVia system leading to a false positive result. Our study demonstrated concordance results between the BD FACSVia system and BD FACSCalibur. Intersite reproducibility of BD HLA-B27 assay remained within the limits of acceptability. © 2018 The Authors. Cytometry Part B: Clinical Cytometry published by Wiley Periodicals, Inc. on behalf of International Clinical Cytometry Society.
BD FACSVia™ 系统是一种具有改进工作流程效率的新型流式细胞仪。为了评估在 BD FACSVia 系统上开发的用于 HLA-B27 的应用程序,我们在三个中心进行了一项一致性研究,以与 BD FACSCalibur™ 系统进行比较。对两个系统均进行了 HLA-B27 检测,对 594 份准备好的供体样本进行了分析。对于 HLA-B27 检测结果不一致的样本,采用反向序列特异性寡核苷酸 (rSSO) DNA 分型方法(LABType® SSO,One Lambda)进行裁决。在 BD FACSVia 系统上观察到 80 个 B27 阳性、499 个 B27 阴性和 15 个“不确定”样本。BD FACSCalibur 系统的相应结果为 73 个 B27 阳性、502 个 B27 阴性和 19 个“灰色区域”样本。两个系统之间 HLA-B27 测定的总体一致性为 98%,与 BD FACSCalibur 相比,BD FACSVia 系统检测到的阳性样本多 7 个。BD FACSVia(命名为“不确定”)和 BD FACSCalibur(命名为“灰色区域”)上阳性和阴性之间的不确定区域是由于 HLA-B27 克隆 GS145.2 的抗体交叉反应所致。一个经 rSSO DNA 方法验证的阴性样本被 BD FACSVia 系统报告为 HLA-B27 阳性,导致假阳性结果。我们的研究表明 BD FACSVia 系统与 BD FACSCalibur 之间存在一致性结果。BD HLA-B27 检测的站点间重现性仍在可接受范围内。© 2018 作者。Wiley Periodicals, Inc. 代表国际临床细胞化学学会出版的《细胞术杂志》B 部分:临床细胞术。