An-Foraker S H, Kawada C Y, McKinney D
Acta Cytol. 1979 Jul-Aug;23(4):303-8.
Endometrial aspiration material obtained with the Isaacs cell sampler from 220 patients was studied and correlated with tissue biopsy from 172 of these patients. Of 52 endometrial carcinomas, there were two false-negative readings. In addition, five cases thought to be malignancies by cytologic criteria were not confirmed by histology. Therefore, the endometrial cell sampler showed a specificity of 96.85%, sensitivity of 96.15% and predictive value of 90.90%. Only nine cases of unsatisfactory cellular samples were obtained as compared to 13 cases of insufficient material by dilatation and curettage.
对使用艾萨克斯细胞采样器从220名患者获取的子宫内膜抽吸物进行了研究,并与其中172名患者的组织活检结果进行了对比。在52例子宫内膜癌病例中,有2例假阴性结果。此外,5例根据细胞学标准被认为是恶性肿瘤的病例未得到组织学证实。因此,子宫内膜细胞采样器的特异性为96.85%,敏感性为96.15%,预测值为90.90%。与刮宫术获取的13例取材不足的病例相比,仅获得9例细胞样本不满意的病例。