Atrial septal defect closure with the new Cardia Ultrasept II™ device with interposed Goretex patch: Mexican experience - has the perforation of Ivalon's membrane been solved?

作者信息

Mijangos-Vázquez Roberto, García-Montes Antonio J, Soto-López Elena M, Guarner-Lans Verónica, Zabal Carlos

机构信息

1Pediatric Interventional Cardiology Department,Pediatric Specialties Hospital,Tuxtla Gutiérrez,Chiapas,México.

2Pediatric Interventional Cardiology Department,National Institute of Cardiology "Ignacio Chávez",Mexico City,Mexico.

出版信息

Cardiol Young. 2018 May;28(5):709-714. doi: 10.1017/S1047951118000100. Epub 2018 Feb 27.

Abstract

OBJECTIVES

The objective of this study was to demonstrate the safety and feasibility of using the new Cardia Ultrasept II™ device with interposed Goretex patch referring to the perforation of polyvinyl alcohol membrane.

BACKGROUND

Great advances have been made in the development of devices for closure of atrial septal defect. The Cardia Ultrasept II™ with interposed Goretex patch is the modified last generation of Cardia devices, having the advantage of a super-low profile within the atria and an integral locking delivery-retrieval mechanism that ensures safe deployment. In addition, with the interposition of the Goretex, it has been possible to abolish perforation of Ivalon's membrane as a complication.Methods and resultsPatients with ostium secundum atrial septal defect with surrounding rims with a minimum length of 5 mm and who underwent atrial septal defect closure with the new Ultrasept II™ with Goretex patch were included from two paediatric cardiac centres. Primary end point was to determine perforation of the Goretex membrane at follow-up; secondary end point included right ventricular diastolic diameter. In total, 30 patients underwent atrial septal defect closure at a median age of 6 (1-29) years. At follow-up for 6 (range, 1-15) months, freedom from perforations was 100%. A continuous decrease in right ventricular diastolic diameter was found with an initial median of 30 (25-49) mm and after catheterisation of 27.5 (18-33) mm, p=0.01, and Z-score of 2.6 (1.7-3.6) versus 1.9 (1-2.9) after procedure, p=0.01.

CONCLUSIONS

The new modified generation of the Ultrasept II™ device with interposed Goretex patch is a good alternative to achieve atrial septal defect closure safely and feasibly with no membrane perforation at follow-up.