Suppr超能文献

一种用于手术室的多普勒流量计。

A Doppler flowmeter for use in theatre.

作者信息

Beard J D, Evans J M, Skidmore R, Horrocks M

出版信息

Ultrasound Med Biol. 1986 Nov;12(11):883-9. doi: 10.1016/0301-5629(86)90006-2.

Abstract

We have developed a Doppler flowmeter based on a 10 MHz pencil probe and mean frequency estimator which overcomes many of the limitations of existing electromagnetic and ultrasonic flowmeters. The output of the flowmeter, which is proportional to the first moment of the Doppler power spectrum and hence mean blood velocity is linear from 1.3 to over 50 cm s-1 for pulsatile flow. Variation in vessel diameter and angle of insonation, which are the common sources of error in Doppler flowmetry, are minimised by constraining the vessel in a plastic cuff which fixes the probe angle at 50 degrees. A simple gauge is used to compress the vessel flat, before the cuff is applied, to measure the wall thickness to within 0.25 mm. The vessel internal diameter and hence blood flow can then be calculated using an experimentally determined calibration factor to compensate for non-even insonation. A range of sterilizable cuffs from 3-12 mm diameter have been built and the flowmeter is now being used routinely during all arterial reconstructive surgery. The accuracy and reproducibility of the system was tested for range of different sized silastic tubes on a hydraulic model and found to be less than 12% for vessels greater than 2 mm internal diameter. Satisfactory signals were easily obtained from all prosthetic materials with the exception of PTFE. The instantaneous output was compared to an electromagnetic flowmeter using a fast Fourier transform algorithm; the moduli of the harmonics were virtually identical but the Doppler system produced a smaller phase shift with increasing harmonics.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

我们已开发出一种基于10兆赫笔式探头和平均频率估计器的多普勒流量计,它克服了现有电磁和超声流量计的许多局限性。该流量计的输出与多普勒功率谱的一阶矩成正比,因此平均血流速度在脉动血流时从1.3至超过50厘米/秒呈线性关系。在多普勒血流测量中常见的误差来源——血管直径和超声入射角的变化,通过将血管置于塑料袖带中得以最小化,该袖带将探头角度固定在50度。在应用袖带之前,使用一个简单的量具将血管压平,以测量壁厚至0.25毫米以内。然后可以使用实验确定的校准因子来计算血管内径,进而计算血流,以补偿不均匀的超声照射。现已制作了一系列直径从3至12毫米的可消毒袖带,该流量计目前已在所有动脉重建手术中常规使用。在液压模型上对不同尺寸的硅橡胶管测试了该系统的准确性和可重复性,发现对于内径大于2毫米的血管,误差小于12%。除了聚四氟乙烯外,从所有假体材料中都很容易获得满意的信号。使用快速傅里叶变换算法将瞬时输出与电磁流量计进行比较;谐波的模量几乎相同,但多普勒系统随着谐波增加产生的相位偏移较小。(摘要截于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验