缩短治疗时间可减少分次内运动的影响:一项比较 VMAT 与前列腺癌分步射野调强放疗的实时 4DUS 研究。

Shorter treatment times reduce the impact of intra-fractional motion : A real-time 4DUS study comparing VMAT vs. step-and-shoot IMRT for prostate cancer.

机构信息

Department of Radiation Oncology, University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany.

出版信息

Strahlenther Onkol. 2018 Jul;194(7):664-674. doi: 10.1007/s00066-018-1286-2. Epub 2018 Mar 9.

Abstract

PURPOSE

To evaluate the impact of shorter treatment times on intra-fractional motion of the prostate during external beam radiotherapy.

METHODS

53 h of intra-fractional motion of the prostate were recorded in real-time by 4D ultrasound (4DUS) during 720 fractions in 28 patients, 14 of which whom treated with step-and-shoot intensity-modulated radiotherapy (IMRT) and 14 of whom were treated with volumetric arc therapy (VMAT).

RESULTS

The average VMAT fraction was recorded for 2 min 43 s and was substantially shorter than the average step-and-shoot IMRT fraction at 6 min 13 s. Average radial displacement of the prostate per fraction was substantially and significantly reduced from 1.31 ± 1.28 mm (n = 357 step-and-shoot IMRT fractions) to 0.96 ± 1.04 mm (n = 363 VMAT fractions), p = 0.00004. Radial, vertical, and longitudinal root-mean-square (r. m. s.) error per fraction was reduced from 1.55 to 1.12 mm (-28%, p < 0.0001), from 1.16 to 0.77 mm (-34%, p < 0.0001), and from 0.79 to 0.56 mm (-29%, p = 0.0002), respectively. Lateral intra-fractional motion was generally small and did not differ significantly. The prostate remained during 95% of fraction time within 4.55 mm of the isocenter in case of step-and-shoot IMRT and within 2.45 mm in case of VMAT. The variance of displacements increased linearly with time, and the rate was the same for both step-and-shoot IMRT and VMAT patients.

CONCLUSIONS

The position of the prostate changed less during shorter fractions, limiting fraction-average and end-of-fraction variance. This substantially and significantly reduced the impact of intra-fractional motion during shorter VMAT fractions as compared to longer step-and-shoot IMRT fractions.

摘要

目的

评估缩短治疗时间对体外放射治疗中前列腺内部分运动的影响。

方法

通过 4D 超声(4DUS)实时记录 28 名患者的 720 个分次中的前列腺内部分运动,共记录 53 小时。其中 14 例采用分步射击强度调制放疗(IMRT)治疗,14 例采用容积弧形治疗(VMAT)治疗。

结果

VMAT 平均分次时间为 2 分 43 秒,明显短于分步射击 IMRT 平均分次时间 6 分 13 秒。前列腺的平均径向位移从 1.31±1.28mm(n=357 个分步射击 IMRT 分次)显著降低至 0.96±1.04mm(n=363 个 VMAT 分次),p=0.00004。每分次的径向、垂直和纵向均方根(r.m.s.)误差分别从 1.55mm 降低至 1.12mm(减少 28%,p<0.0001)、从 1.16mm 降低至 0.77mm(减少 34%,p<0.0001)、从 0.79mm 降低至 0.56mm(减少 29%,p=0.0002)。侧向的分次内运动通常较小,差异无统计学意义。在分步射击 IMRT 的情况下,前列腺在 95%的分次时间内保持在等中心 4.55mm 以内,在 VMAT 的情况下保持在 2.45mm 以内。位移的方差随时间线性增加,分步射击 IMRT 和 VMAT 患者的速率相同。

结论

较短的分次中前列腺的位置变化较小,限制了分次平均和分次末方差。与较长的分步射击 IMRT 分次相比,这显著降低了较短 VMAT 分次中内部分运动的影响。