Changes in Optical Density of Postlens Fluid Reservoir During 2 Hours of Scleral Lens Wear.

作者信息

Schornack Muriel M, Nau Cherie B

机构信息

Department of Ophthalmology, Mayo Clinic, Rochester, MN.

出版信息

Eye Contact Lens. 2018 Nov;44 Suppl 2:S344-S349. doi: 10.1097/ICL.0000000000000500.

Abstract

OBJECTIVES

Debris accumulation in the postlens fluid reservoir during scleral lens wear is clinically observed. We evaluated a method to quantify increases in turbidity of the postlens fluid reservoir by assessing changes in optical density of the fluid over time and compared these changes during 2 hr of scleral lens wear using three different lens designs.

METHODS

Thirty-five patients (age, mean [SD], 29 [7] years) with no history of eye disease were enrolled in the study. Participants were fit with a 15.0-, 18.0-, and 18.2-mm scleral lens on one randomly selected eye during the enrollment examination. During each of three subsequent visits, one of three lenses selected during enrollment was placed on the eye. Scheimpflug images were acquired within 5 min of lens placement and at 20-min intervals for the ensuing 2 hr. Fluid reservoir optical density was assessed using Pentacam (Oculus Inc) analysis software.

RESULTS

Increases in fluid reservoir optical density were significant between each time interval with all lenses until 100 min. No significant increases in fluid reservoir optical density were noted with the 15.0- and 18.0-mm lenses between 100 and 120 min; fluid reservoir optical density increased with the 18.2-mm lens between 100 and 120 min. Optical density increased by 105% (15.0-mm lens) and 117% (18.0- and 18.2-mm lenses) at 2 hr.

CONCLUSIONS

Optical density of the postlens fluid reservoir can be quantified during scleral lens wear. During 2 hr of wear, fluid reservoir optical density doubled with all lens designs evaluated.

摘要

目的

临床上观察到巩膜镜佩戴期间晶状体后房液体储存器中有碎屑堆积。我们评估了一种通过评估液体光密度随时间的变化来量化晶状体后房液体储存器浑浊度增加的方法,并比较了使用三种不同镜片设计在巩膜镜佩戴2小时期间的这些变化。

方法

35名无眼部疾病史的患者(年龄,平均[标准差],29[7]岁)参与了该研究。在入组检查期间,为参与者一只随机选择的眼睛佩戴15.0、18.0和18.2毫米的巩膜镜。在随后的三次就诊中,每次将入组时选择的三种镜片之一放置在眼睛上。在镜片放置后5分钟内以及随后2小时内每隔20分钟采集一次Scheimpflug图像。使用Pentacam(奥culus公司)分析软件评估液体储存器的光密度。

结果

在100分钟之前,所有镜片在每个时间间隔之间液体储存器光密度均显著增加。15.0毫米和18.0毫米镜片在100至120分钟之间液体储存器光密度无显著增加;18.2毫米镜片在100至120分钟之间液体储存器光密度增加。2小时时光密度增加了105%(15.0毫米镜片)和117%(18.0毫米和18.2毫米镜片)。

结论

巩膜镜佩戴期间可量化晶状体后房液体储存器的光密度。在佩戴2小时期间,所有评估的镜片设计中液体储存器光密度均增加了一倍。