在一个短期研究环境中,对74只接受起搏器植入的犬进行的146次主动和被动固定心动过缓导线植入术进行回顾性研究。

A retrospective review of 146 active and passive fixation bradycardia lead implantations in 74 dogs undergoing pacemaker implantation in a research setting of short term duration.

作者信息

Swanson Lynne E, Huibregtse Barbara A, Scansen Brian A

机构信息

Research and Technology Center, Boston Scientific Corporation, 4100 Hamline Ave North, Saint Paul, MN, 55112, USA.

Preclinical Sciences, Boston Scientific Corporation, 100 Boston Scientific Way, Marlborough, MA, 01752, USA.

出版信息

BMC Vet Res. 2018 Mar 27;14(1):112. doi: 10.1186/s12917-018-1431-2.

Abstract

BACKGROUND

Canine veterinary patients increasingly benefit from implantation of transvenous pacemakers for bradyarrhythmias. No published data exist examining procedural outcomes of pacemaker implantation performed in the preclinical laboratory. The purpose was to review short term complication, infection, dislodgement, penetration rates, plus overall morbidity following pacemaker implantation in the research setting. A retrospective review of 74 Class A purpose-bred mongrels implanted with active (n = 89) and passive fixation (n = 57) intracardiac leads for dual (n = 72) or single (n = 2) chamber pacing was performed.

RESULTS

All leads were implanted successfully, meeting electrical implant criteria. Follow-ups typically occurred every 7 days (first month), then at 30 day intervals. Seroma formation was 1.4% and 10.8% at the venotomy and pulse generator site respectively. Overall infection rate was 1.4%. Overall dislodgement rate was 2.1%, (2 passive atrial leads, 1 passive ventricular lead). Overall fractures and insulation defects were zero. Two helix penetrations were noted incidentally post mortem, one at the right atrial appendage and one at the right ventricle (64 dogs, 128 leads evaluated), a 1.6% event rate. Major in-life adverse events were 5.4% (4 of 74 dogs), including 1 infection and 3 lead dislodgements.

CONCLUSIONS

This review demonstrates a low complication rate with bradycardia lead implants in the short term (up to 180 days), in a high volume research setting. Lead type, implant technique, surgeon experience, healthy patient population, patient size and follow-up care play a role. This review also suggests active fixation leads in the right atrial appendage of dogs are safe and reliable.

摘要

背景

犬类兽医患者越来越多地受益于经静脉起搏器植入以治疗缓慢性心律失常。目前尚无已发表的数据研究在临床前实验室进行的起搏器植入手术的结果。目的是回顾在研究环境中起搏器植入后的短期并发症、感染、脱位、穿透率以及总体发病率。对74只A类目的繁育杂种犬进行了回顾性研究,这些犬植入了主动(n = 89)和被动固定(n = 57)的心内导联用于双腔(n = 72)或单腔(n = 2)起搏。

结果

所有导联均成功植入,符合电植入标准。随访通常每7天(第一个月)进行一次,然后每隔30天进行一次。静脉切开术和脉冲发生器部位的血清肿形成率分别为1.4%和10.8%。总体感染率为1.4%。总体脱位率为2.1%(2根被动心房导联,1根被动心室导联)。总体骨折和绝缘缺陷率为零。尸检时偶然发现两例螺旋穿透,一例位于右心耳,一例位于右心室(64只犬,评估128根导联),发生率为1.6%。主要的术中不良事件发生率为5.4%(74只犬中的4只),包括1例感染和3例导联脱位。

结论

本综述表明,在高容量研究环境中,短期内(长达180天)缓慢性心律失常导联植入的并发症发生率较低。导联类型、植入技术、外科医生经验、健康的患者群体、患者体型和随访护理都起到了一定作用。本综述还表明,在犬的右心耳中使用主动固定导联是安全可靠的。