自动光谱成像协议选择与迭代重建相结合可提高腹部CT血管造影的图像质量,并减少辐射和造影剂用量。

Automatic spectral imaging protocol selection combined with iterative reconstruction can enhance image quality and decrease radiation and contrast dosage in abdominal CT angiography.

作者信息

Yin Xiao-Ping, Gao Bu-Lang, Li Cai-Ying, Zuo Zi-Wei, Xu Ying-Jin, Wang Jia-Ning, Liu Huai-Jun, Liang Guang-Lu

机构信息

CT/MRI Room, Affiliated Hospital of Hebei University, Baoding, 071000, Hebei, People's Republic of China.

Department of Medical Research, Shijiazhuang First Hospital, Shijiazhuang, 050011, Hebei, People's Republic of China.

出版信息

Jpn J Radiol. 2018 May;36(5):345-350. doi: 10.1007/s11604-018-0734-3. Epub 2018 Apr 3.

Abstract

PURPOSE

To investigate the effect of automatic spectral imaging protocol selection (ASIS) and adaptive statistical iterative reconstruction (ASIR) technology in reducing radiation and contrast dosage.

METHODS

Sixty-four patients were randomly divided into two groups for abdominal computed tomography (CT): the experiment group with ASIS plus 50% ASIR and the control with 120 kVp voltage.

RESULTS

The CT dose-index volume decreased by 23.68 and 23.57% and the dose-length product dropped by 25.59 and 18.45% in the arterial and portal venous phases, respectively, in the experiment than control group. The contrast dose was reduced by 16.86% in the experiment group. In the 55 keV + 50% ASIR group, the arterial contrast-to-noise ratio and scores were significantly (P < 0.05) higher than in the control group in the arterial phase while the portal contrast-to-noise ratio and scores were not significantly different between the two groups (P > 0.05).

CONCLUSION

The ASIS technique plus 50% ASIR can enhance image quality of the abdominal structures while decreasing the radiation and contrast dosage compared with the conventional scan mode.

摘要

目的

探讨自动光谱成像协议选择(ASIS)和自适应统计迭代重建(ASIR)技术在降低辐射剂量和对比剂剂量方面的效果。

方法

将64例患者随机分为两组进行腹部计算机断层扫描(CT):实验组采用ASIS加50%ASIR,对照组采用120 kVp电压。

结果

与对照组相比,实验组动脉期和门静脉期的CT剂量指数容积分别降低了23.68%和23.57%,剂量长度乘积分别降低了25.59%和18.45%。实验组的对比剂剂量降低了16.86%。在55 keV + 50%ASIR组中,动脉期的动脉对比噪声比和评分显著高于对照组(P < 0.05),而门静脉期两组的对比噪声比和评分无显著差异(P > 0.05)。

结论

与传统扫描模式相比,ASIS技术加50%ASIR可提高腹部结构的图像质量,同时降低辐射剂量和对比剂剂量。