Yin Xiao-Ping, Gao Bu-Lang, Li Cai-Ying, Zuo Zi-Wei, Xu Ying-Jin, Wang Jia-Ning, Liu Huai-Jun, Liang Guang-Lu
CT/MRI Room, Affiliated Hospital of Hebei University, Baoding, 071000, Hebei, People's Republic of China.
Department of Medical Research, Shijiazhuang First Hospital, Shijiazhuang, 050011, Hebei, People's Republic of China.
Jpn J Radiol. 2018 May;36(5):345-350. doi: 10.1007/s11604-018-0734-3. Epub 2018 Apr 3.
To investigate the effect of automatic spectral imaging protocol selection (ASIS) and adaptive statistical iterative reconstruction (ASIR) technology in reducing radiation and contrast dosage.
Sixty-four patients were randomly divided into two groups for abdominal computed tomography (CT): the experiment group with ASIS plus 50% ASIR and the control with 120 kVp voltage.
The CT dose-index volume decreased by 23.68 and 23.57% and the dose-length product dropped by 25.59 and 18.45% in the arterial and portal venous phases, respectively, in the experiment than control group. The contrast dose was reduced by 16.86% in the experiment group. In the 55 keV + 50% ASIR group, the arterial contrast-to-noise ratio and scores were significantly (P < 0.05) higher than in the control group in the arterial phase while the portal contrast-to-noise ratio and scores were not significantly different between the two groups (P > 0.05).
The ASIS technique plus 50% ASIR can enhance image quality of the abdominal structures while decreasing the radiation and contrast dosage compared with the conventional scan mode.
探讨自动光谱成像协议选择(ASIS)和自适应统计迭代重建(ASIR)技术在降低辐射剂量和对比剂剂量方面的效果。
将64例患者随机分为两组进行腹部计算机断层扫描(CT):实验组采用ASIS加50%ASIR,对照组采用120 kVp电压。
与对照组相比,实验组动脉期和门静脉期的CT剂量指数容积分别降低了23.68%和23.57%,剂量长度乘积分别降低了25.59%和18.45%。实验组的对比剂剂量降低了16.86%。在55 keV + 50%ASIR组中,动脉期的动脉对比噪声比和评分显著高于对照组(P < 0.05),而门静脉期两组的对比噪声比和评分无显著差异(P > 0.05)。
与传统扫描模式相比,ASIS技术加50%ASIR可提高腹部结构的图像质量,同时降低辐射剂量和对比剂剂量。