Suppr超能文献

手动加载和预加载的可折叠丙烯酸人工晶状体的扫描电子显微镜特征

Scanning electron microscopic characteristics of manually loaded and preloaded foldable acrylic intraocular lenses.

作者信息

Rateb Mahmoud Fathy, Eldaly Zeiad Hasan, Ibrahim Walid Saad Eldein, Eldoghaimy Ahmed Hasan

机构信息

1 Department of Ophthalmology, Assiut University Hospitals, Assiut, Egypt.

2 Ophthalmology Department, South Valley University Hospital, Qena, Egypt.

出版信息

Eur J Ophthalmol. 2019 Jan;29(1):28-32. doi: 10.1177/1120672118762665. Epub 2018 Apr 5.

Abstract

PURPOSE

: To compare surface alterations between preloaded and manually loaded intraocular lens.

METHODS

: Scanning electron microscope was utilized to evaluate surface alteration and deposits in four different types of intraocular lenses: preloaded hydrophobic acrylic, preloaded hydrophilic acrylic, manually loaded hydrophobic acrylic, and manually loaded hydrophilic acrylic. Six lenses with different powers (+6 D, +22 D, and +29 D) were used from each category, to represent different thickness categories of the intraocular lenses.

RESULTS

: In total, 30 intraocular lenses have been evaluated in this study: 4 from the control group (2 hydrophobic and 2 hydrophilic lenses) and 12 from the preloaded intraocular lens and manually loaded groups (6 hydrophilic and 6 hydrophobic lenses with different powers). Surface deposits were found in eight hydrophobic intraocular lenses compared to a single intraocular lens with scattered deposits on the optical surface of a hydrophilic intraocular lens. In manually loaded intraocular lens group, five hydrophobic and one hydrophilic intraocular lenses showed identifiable marks on the optical surface. In the preloaded intraocular lens group, three hydrophobic intraocular lenses showed identifiable marks on the optical surface and three hydrophobic intraocular lenses showed surface wrinkling. All hydrophilic intraocular lenses revealed no identifiable marks.

CONCLUSION

: Surface alterations and deposits are a common finding in both preloaded and manually loaded intraocular lenses. Water content of acrylic intraocular lenses is an important factor predisposing to these changes, and hydrophobic intraocular lenses are more vulnerable than hydrophilic lenses. The impact on the final visual outcome needs further studies.

摘要

目的

比较预装式和手动植入式人工晶状体的表面变化。

方法

利用扫描电子显微镜评估四种不同类型人工晶状体的表面变化和沉积物:预装式疏水丙烯酸酯人工晶状体、预装式亲水丙烯酸酯人工晶状体、手动植入式疏水丙烯酸酯人工晶状体和手动植入式亲水丙烯酸酯人工晶状体。从每个类别中使用六个不同屈光度(+6 D、+22 D和+29 D)的晶状体,以代表人工晶状体的不同厚度类别。

结果

本研究共评估了30个人工晶状体:对照组4个(2个疏水型和2个亲水型晶状体),预装式人工晶状体组和手动植入式组各12个(6个亲水型和6个疏水型晶状体,屈光度不同)。与一个在亲水型人工晶状体光学表面有散在沉积物的人工晶状体相比,在八个疏水型人工晶状体中发现了表面沉积物。在手动植入式人工晶状体组中,五个疏水型和一个亲水型人工晶状体在光学表面有可识别的痕迹。在预装式人工晶状体组中,三个疏水型人工晶状体在光学表面有可识别的痕迹,三个疏水型人工晶状体有表面起皱现象。所有亲水型人工晶状体均未发现可识别的痕迹。

结论

表面变化和沉积物在预装式和手动植入式人工晶状体中均很常见。丙烯酸酯人工晶状体的含水量是导致这些变化的一个重要因素,疏水型人工晶状体比亲水型人工晶状体更容易出现这些变化。对最终视觉效果的影响需要进一步研究。