Jang Hyunsik, Hur Jin, Choi Byoung Wook, Im Dong Jin, Hong Yoo Jin, Kim Young Jin, Lee Hye-Jeong
Department of Radiology, Severance Hospital, Research Institute of Radiological Science, Yonsei University, College of Medicine, Seoul, Republic of Korea.
Acta Radiol. 2018 Dec;59(12):1475-1481. doi: 10.1177/0284185118768098. Epub 2018 Apr 10.
Although the bismuth breast shield can reduce radiation exposure to the breast during dual-energy computed tomography (DECT), it can potentially affect material quantification on DECT due to artifacts.
To evaluate the effects of bismuth breast shielding on iodine quantification and radiation exposure in DECT.
Small balloons were made with 0.2%, 0.6% and 1.0% blended iodinated contrast (370 mg/mL of iodine) with water. The balloons were located at both anterior and posterior lungs in an adult anthropomorphic chest phantom. DECT was performed with and without breast shielding. Afterwards, iodine concentration values were measured for each balloon on the iodine maps. Absorbed radiation doses in the breast were measured with the optically stimulated luminescence dosimeter.
After shielding, we obtained significantly decreased iodine quantification for all three concentrations with 0.78 ± 0.13 to 0.46 ± 0.13 mg/mL, 2.31 ± 0.17 to 1.68 ± 0.19 mg/mL, and 3.82 ± 0.10 to 2.84 ± 0.20 mg/mL at the anterior location, and 0.72 ± 0.11 to 0.48 ± 0.09 mg/mL, 2.24 ± 0.13 to 1.87 ± 0.21 mg/mL, and 3.75 ± 0.16 to 3.15 ± 0.14 mg/mL at the posterior location for the 0.2%, 0.6%, and 1.0% balloons, respectively ( P = 0.001 for all). After shielding, absorbed radiation doses to the breast significantly decreased by 14.8% (4.32 ± 0.33 to 3.68 ± 0.30 mGy; P = 0.005).
Although using the bismuth breast shield may decrease radiation exposure to the breast on DECT, it may also significantly affect iodine quantification.