Suppr超能文献

应用和优势单能量重建图像减少金属伪影使用双能 CT 膝关节和髋关节假体。

Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses.

机构信息

Institute of Radiology, Catholic University School of Medicine, Fondazione Policlinico Universitario A. Gemelli, Rome, Italy.

Institute of Clinical Orthopedic, Catholic University School of Medicine, Fondazione Policlinico Universitario A. Gemelli, Rome, Italy.

出版信息

Radiol Med. 2018 Aug;123(8):593-600. doi: 10.1007/s11547-018-0881-8. Epub 2018 Apr 10.

Abstract

OBJECTIVE

The study aimed to assess image quality when using dual-energy CT (DECT) to reduce metal artifacts in subjects with knee and hip prostheses.

METHODS

Twenty-two knee and 10 hip prostheses were examined in 31 patients using a DECT protocol (tube voltages 100 and 140 kVp). Monoenergetic reconstructions were extrapolated at 64, 69, 88, 105, 110, 120, 140, 170, and 190 kilo-electron volts (keV) and the optimal energy was manually selected. The B60-140 and Fast DE reconstructions were made by CT. The image quality and diagnostic value were subjectively and objectively determined. Double-blind qualitative assessment was performed by two radiologists using a Likert scale. For quantitative analysis, a circular region of interest (ROI) was placed by a third radiologist within the most evident streak artifacts on every image. Another ROI was placed in surrounding tissues without artifacts as a reference.

RESULTS

The inter-reader agreement for the qualitative assessment was nearly 100%. The best overall image quality (37.8% rated "excellent") was the Fast DE Siemens reconstruction, followed by B60-140 and Opt KeV (20.5 and 10.2% rated excellent). On the other hand, DECT images at 64, 69 and 88 keV had the worse scores. The number of artifacts was significantly different between monoenergetic images. Nevertheless, because of the high number of pairwise comparisons, no differences were found in the post hoc analysis except for a trend toward statistical significance when comparing the 170 and 64 keV doses.

CONCLUSIONS

DECT with specific post-processing may reduce metal artifacts and significantly enhance the image quality and diagnostic value when evaluating metallic implants.

摘要

目的

本研究旨在评估使用双能 CT(DECT)减少膝关节和髋关节假体金属伪影的图像质量。

方法

31 例患者共 31 个膝关节和 10 个髋关节假体接受 DECT 方案(管电压 100 和 140kVp)检查。在 64、69、88、105、110、120、140、170 和 190 千电子伏特(keV)处推断出单能量重建,并手动选择最佳能量。B60-140 和 Fast DE 重建由 CT 完成。主观和客观评估图像质量和诊断价值。两位放射科医生使用李克特量表进行双盲定性评估。为了进行定量分析,第三位放射科医生在每张图像上最明显的条纹伪影内放置一个圆形感兴趣区(ROI)。另一个 ROI 放置在无伪影的周围组织中作为参考。

结果

定性评估的两位读者间一致性接近 100%。最佳整体图像质量(37.8%评为“优秀”)是 Fast DE Siemens 重建,其次是 B60-140 和 Opt KeV(分别有 20.5%和 10.2%评为优秀)。另一方面,64keV、69keV 和 88keV 的 DECT 图像评分最差。单能量图像之间的伪影数量存在显著差异。然而,由于进行了大量的成对比较,除了在比较 170keV 和 64keV 剂量时存在统计学意义的趋势外,在事后分析中未发现差异。

结论

使用特定的后处理技术的 DECT 可以减少金属伪影,并在评估金属植入物时显著提高图像质量和诊断价值。