换能器类型对骨导听力阈值的影响。

Influence of transducer types on bone conduction hearing thresholds.

机构信息

University Hospital Halle (Saale), Department of Otorhinolaryngology, Head and Neck Surgery, Martin Luther University Halle-Wittenberg, Halle, Germany.

出版信息

PLoS One. 2018 Apr 11;13(4):e0195233. doi: 10.1371/journal.pone.0195233. eCollection 2018.

Abstract

OBJECTIVE

Different types of bone conduction transducers with different physical and electro-acoustic properties are available for audiometric hearing threshold measurements. The reference equivalent threshold vibratory force levels (RETVFL) specified in ISO 389-3 are based on measurements conducted with the B71 and KH70 transducers but apply to all types of transducers available for bone conduction audiometry. The objective of this study was to compare bone conduction hearing thresholds measured by different transducers.

DESIGN

In a prospective study the hearing thresholds were measured psychometrically between 125 Hz and 8000 Hz using the Radioear B71, B81 and Präcitronic KH70 transducers.

STUDY SAMPLE

Twenty-one normal hearing participants and fifteen hearing impaired participants.

RESULTS

In both groups significant differences were found between the thresholds measured with the different transducers at the low frequencies 125 Hz and 250 Hz and the high frequencies 3000 Hz, 4000 Hz, 6000 Hz and 8000 Hz. In the normal hearing group, deviations from the reference threshold 0 dB HL towards lower thresholds were observed for the B71 and B81 at 125 Hz and at the high frequencies 3000 Hz, 4000 Hz, 6000 Hz and 8000 Hz.

CONCLUSIONS

RETVFL-values should be reassessed and provided specifically for the different transducers.

摘要

目的

不同类型的骨导换能器具有不同的物理和电声特性,可用于测听听力阈值。ISO 389-3 中规定的参考等效阈振动力级(RETVFL)基于使用 B71 和 KH70 换能器进行的测量,但适用于所有可用于骨导测听的换能器类型。本研究的目的是比较不同换能器测量的骨导听力阈值。

设计

在一项前瞻性研究中,使用 Radioear B71、B81 和 Präcitronic KH70 换能器,在 125 Hz 至 8000 Hz 之间进行心理物理听力阈值测量。

研究样本

21 名正常听力参与者和 15 名听力受损参与者。

结果

在两组中,在低频 125 Hz 和 250 Hz 以及高频 3000 Hz、4000 Hz、6000 Hz 和 8000 Hz 处,使用不同换能器测量的阈值之间存在显著差异。在正常听力组中,在低频 125 Hz 和高频 3000 Hz、4000 Hz、6000 Hz 和 8000 Hz 处,B71 和 B81 的参考阈值 0 dB HL 向较低阈值的偏差。

结论

应重新评估 RETVFL 值,并为不同的换能器提供具体值。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f4cf/5895012/0f3eb02b3269/pone.0195233.g001.jpg

相似文献

1
Influence of transducer types on bone conduction hearing thresholds.换能器类型对骨导听力阈值的影响。
PLoS One. 2018 Apr 11;13(4):e0195233. doi: 10.1371/journal.pone.0195233. eCollection 2018.

引用本文的文献

本文引用的文献

8
Acoustic radiation from bone vibrators.骨振动器的声辐射。
Ear Hear. 1981 Mar-Apr;2(2):59-63. doi: 10.1097/00003446-198103000-00001.
10
Intertest variability and the air-bone gap.
J Speech Hear Disord. 1967 Feb;32(1):82-6. doi: 10.1044/jshd.3201.82.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验