This proceedings report presents the outcomes from an international Expert Meeting to establish a consensus on the recommended technical and operational requirements for air quality within modern assisted reproduction technology (ART) laboratories. Topics considered included design and construction of the facility, as well as its heating, ventilation and air conditioning system; control of particulates, micro-organisms (bacteria, fungi and viruses) and volatile organic compounds (VOCs) within critical areas; safe cleaning practices; operational practices to optimize air quality while minimizing physicochemical risks to gametes and embryos (temperature control versus air flow); and appropriate infection-control practices that minimize exposure to VOC. More than 50 consensus points were established under the general headings of assessing site suitability, basic design criteria for new construction, and laboratory commissioning and ongoing VOC management. These consensus points should be considered as aspirational benchmarks for existing ART laboratories, and as guidelines for the construction of new ART laboratories.
本会议报告呈现了一次国际专家会议的成果,旨在就现代辅助生殖技术 (ART) 实验室空气质量的推荐技术和操作要求达成共识。所审议的议题包括设施的设计和建造,以及其供暖、通风和空调系统;关键区域内颗粒物、微生物(细菌、真菌和病毒)和挥发性有机化合物 (VOC) 的控制;安全清洁实践;优化空气质量的操作实践,同时将配子和胚胎的理化风险降至最低(温度控制与气流);以及最大限度减少接触 VOC 的适当感染控制实践。在评估场地适宜性、新建筑的基本设计标准以及实验室调试和持续的 VOC 管理等一般标题下,确立了 50 多个共识点。这些共识点应被视为现有 ART 实验室的理想基准,也是新建 ART 实验室的指南。