Suppr超能文献

在使用 Surpass 装置进行血流导向栓塞治疗时,将中间导管置于脑动脉瘤的远端。

Intermediate catheter placement distal to the cerebral aneurysm during flow diversion embolization with the Surpass device.

机构信息

Department of Radiology, Yeditepe University Hospital, Istanbul, Turkey.

Department of Radiology, Hacettepe University Hospitals, Ankara, Turkey.

出版信息

J Neurointerv Surg. 2018 Dec;10(12):e35. doi: 10.1136/neurintsurg-2017-013662. Epub 2018 Apr 25.

Abstract

PURPOSE

We present the preliminary angiographic and clinical results of the combined utilization of an AXS Catalyst 5 (Cat5) distal access catheter and a Surpass flow diverter (SFD) for the endovascular treatment of intracranial aneurysms.

MATERIALS AND METHODS

Interventional neuroradiology case records were evaluated retrospectively to identify patients treated with an SFD in combination with Cat5. Demographic data, technical success, location of the catheter tip, aneurysm occlusion rate, and procedure-related morbidity and mortality were noted.

RESULTS

The analysis yielded 24 patients with 25 aneurysms with a mean age of 52.3±19.7 years. The mean aneurysm diameter was 15.7±8.8 mm. Twenty-one were anterior circulation (seven supraclinoid, one middle cerebral, one anterior choroidal and 12 petrocavernous) and four were posterior circulation (three basilar tip, one posterior inferior cerebellar) aneurysms. Only two of these aneurysms were re-treatments for recurrences. For SFD deployment, the distal tip of Cat5 was placed intradurally and at, or distal to, M1 segment in 88% and 48% of patients, respectively. Aneurysms were bypassed with Cat5 in 80% of the procedures. In one patient, flow diversion was abandoned totally due to arterial tortuosity. With the exclusion of this patient, the technical success rate was 91.7%. There was no permanent procedure-related morbidity or mortality. Of the 22 aneurysms treated successfully with the Surpass device, 20 had imaging follow-up at 6.3±3.8 months and 95% were totally occluded.

CONCLUSION

The Cat5 catheter, which was manufactured for the navigation of large-bore stent-like devices, is a useful adjunct to the Surpass device and allowed us to deploy the device safely.

摘要

目的

我们展示了联合使用 AXS Catalyst 5(Cat5)远端通路导管和 Surpass 血流导向装置(SFD)治疗颅内动脉瘤的初步血管造影和临床结果。

材料和方法

回顾性评估介入神经放射学病例记录,以确定接受 SFD 联合 Cat5 治疗的患者。记录人口统计学数据、技术成功率、导管尖端位置、动脉瘤闭塞率以及与程序相关的发病率和死亡率。

结果

分析得出 24 例患者 25 个动脉瘤,平均年龄 52.3±19.7 岁。平均动脉瘤直径为 15.7±8.8mm。21 个为前循环(7 个前交通,1 个大脑中动脉,1 个前脉络膜和 12 个海绵窦),4 个为后循环(3 个基底动脉尖,1 个小脑后下动脉)动脉瘤。其中只有两个动脉瘤是为复发而进行的再治疗。对于 SFD 的部署,Cat5 的远端尖端分别位于椎管内和 M1 段内或其远侧,在 88%和 48%的患者中。在 80%的手术中,Cat5 绕过了动脉瘤。在一名患者中,由于动脉迂曲,血流转向完全被放弃。排除这名患者后,技术成功率为 91.7%。没有永久性与程序相关的发病率或死亡率。在使用 Surpass 装置成功治疗的 22 个动脉瘤中,有 20 个在 6.3±3.8 个月进行了影像学随访,95%完全闭塞。

结论

Cat5 导管是为大口径支架样装置的导航而制造的,它是 Surpass 装置的有用辅助工具,使我们能够安全地部署该装置。