基于曲面扫描配准的锥形束 CT 扫描中修复体治疗计划的位置精度。

Positional accuracy of a prosthetic treatment plan incorporated into a cone beam computed tomography scan using surface scan registration.

机构信息

Graduate student, Advanced Education in Prosthodontics Program, Division of Restorative Sciences and Prosthodontics, The Ohio State University College of Dentistry, Columbus, Ohio.

Research Program Director and Associate Professor, Division of Orthodontics, The Ohio State University College of Dentistry, Columbus, Ohio.

出版信息

J Prosthet Dent. 2018 Sep;120(3):367-374. doi: 10.1016/j.prosdent.2017.11.019. Epub 2018 Apr 25.

Abstract

STATEMENT OF PROBLEM

The registration of surface scans onto cone beam computed tomography (CBCT) scans has been proposed as a method of visualizing different anatomic structures and the prosthetic treatment plan simultaneously. This method also overcomes some of the problems associated with conventional radiographic templates. However, it has not been thoroughly investigated or validated for use in implant dentistry.

PURPOSE

The purpose of this in vitro study was to evaluate the accuracy of a prosthetic treatment plan surface scan incorporated into a CBCT scan and to compare it to conventional radiographic templates for single tooth replacement.

MATERIAL AND METHODS

Direct surface scans of a completely dentate master model with removable radiopaque teeth were made using an intraoral scanner, and indirect surface scans of a stone duplicate of the master model were made using a laboratory scanner. To simulate a clinical scenario, the mandibular left first molar was removed. A CBCT scan of the clinical scenario was made. The surface scans were registered onto the CBCT scans. Radiographic templates for the clinical scenario were fabricated, and the master model was subsequently scanned using the same CBCT scanner with each radiographic template seated. Metrology software was used to assess the accuracy of each method by measuring the 3-dimensional deviation on standard tessellation language (STL) files generated from the CBCT scans against an STL file of the completely dentate master model generated from a CBCT scan. One-way ANOVA and the Tukey HSD test were used for statistical analysis (α=.05).

RESULTS

The incorporation technique had a significant effect on deviation from the master model (P=.004). The overall mean 3-dimensional deviation was 0.04 mm for direct surface scan registrations, 0.03 mm for indirect surface scan registrations, and 0.33 mm for radiographic templates. Radiographic templates were significantly less accurate compared with both surface scan registration methods (P<.001), whereas no statistically significant difference in accuracy was found between the registration of a laboratory scan and that of an intraoral scan (P=.94).

CONCLUSIONS

Intraoral and laboratory surface scan registration on CBCT scans were more accurate than radiographic templates for prosthetic treatment plan incorporation. The accuracy of the registration of an intraoral scan and a laboratory scan was comparable.

摘要

问题陈述

将表面扫描注册到锥形束计算机断层扫描 (CBCT) 扫描中已被提议作为同时可视化不同解剖结构和修复治疗计划的方法。这种方法还克服了一些与传统放射模板相关的问题。然而,它尚未经过彻底研究或验证,不能用于种植牙。

目的

本体外研究的目的是评估整合到 CBCT 扫描中的修复治疗计划表面扫描的准确性,并将其与用于单颗牙齿置换的传统放射模板进行比较。

材料和方法

使用口腔内扫描仪对带有可移动射线照相牙齿的全牙列主模型进行直接表面扫描,并使用实验室扫描仪对主模型的石质复制品进行间接表面扫描。为了模拟临床情况,去除下颌左侧第一磨牙。对临床情况进行 CBCT 扫描。将表面扫描注册到 CBCT 扫描中。制作临床情况的放射模板,然后用每个放射模板就位的相同 CBCT 扫描仪对主模型进行扫描。使用计量软件通过测量从 CBCT 扫描生成的标准网格语言 (STL) 文件与从 CBCT 扫描生成的全牙列主模型的 STL 文件之间的 3 维偏差来评估每种方法的准确性。使用单因素方差分析和 Tukey HSD 检验进行统计分析 (α=.05)。

结果

整合技术对与主模型的偏差有显著影响 (P=.004)。直接表面扫描注册的总体平均 3 维偏差为 0.04 毫米,间接表面扫描注册的为 0.03 毫米,放射模板的为 0.33 毫米。与两种表面扫描注册方法相比,放射模板的准确性明显较低 (P<.001),而口腔内扫描和实验室扫描的注册精度没有统计学差异 (P=.94)。

结论

CBCT 扫描上的口腔内和实验室表面扫描注册比放射模板更准确,可用于修复治疗计划的整合。口腔内扫描和实验室扫描的注册精度相当。