Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing 100084, China.
Shenzhen Key Laboratory for Minimal Invasive Medical Technologies, Institute of Optical Imaging and Sensing, Graduate School at Shenzhen, Tsinghua University, Shenzhen 518055, China.
Sensors (Basel). 2018 May 14;18(5):1550. doi: 10.3390/s18051550.
A phase-sensitive weak measurement biosensor was proposed for the detection of carcinoembryonic antigen (CEA), one common category of tumor markers. The total internal reflection (TIR) at the interface of the prism without precious metal coating was exploited to introduce the phase delay between horizontal and vertical polarizations, which can be determined through the central wavelength shift of output spectra for the sensing of the refractive index of the sample. In the weak measurement analysis, the specific binding reaction of tumor markers with a refractive index change on the surface of the prism can be monitored in real time through the central wavelength shift. With the specific absorption measurement, the feasibility of this weak measurement-based biosensor was experimentally demonstrated. We provide a low cost and convenient approach for tumor marker detection.
一种用于检测癌胚抗原(CEA)等常见肿瘤标志物的相敏弱测量生物传感器被提了出来。利用没有贵金属涂层的棱镜界面上的全内反射(TIR)引入水平偏振和垂直偏振之间的相位延迟,这可以通过输出光谱的中心波长移动来确定,用于感测样品的折射率。在弱测量分析中,通过中心波长移动可以实时监测肿瘤标志物与棱镜表面折射率变化的特异性结合反应。通过特定的吸收测量,实验验证了这种基于弱测量的生物传感器的可行性。我们为肿瘤标志物的检测提供了一种低成本、方便的方法。